可穿戴无线无创光电容积描记设备与脉搏轮廓心输出量在综合ICU中的心输出量测量比较:简要报告

Comparing Cardiac Output Measurements Using a Wearable, Wireless, Noninvasive Photoplethysmography-Based Device to Pulse Contour Cardiac Output in the General ICU: A Brief Report.
PubMed 文献 · Crit Care Explor · 2022-02-01
一句话结论在7例综合ICU患者69次测量中,胸贴式光电容积描记设备测定的心输出量与脉搏轮廓/热稀释法高度相关(r=0.906),偏差0.3 L/min。

本研究为单中心前瞻性观察性比较临床试验,在综合ICU纳入按主治医师决定使用脉搏轮廓心输出量设备监测的患者,平行使用光电容积描记胸贴设备监测心输出量,医疗团队对无创设备数值设盲。最终分析7例患者共69次测量,热稀释法与光电容积描记法平均心输出量分别为7.3±2.0 L/min和7.0±1.5 L/min,Bland-Altman偏差0.3 L/min(95%一致性界限-1.6至2.2 L/min),百分比偏差2.4%(95%一致性界限-25.7%至30.5%),两法高度相关(r=0.906)。作者认为该胸贴设备与脉搏轮廓心输出量及热稀释法相比具有高准确性,可在多种临床场景提供无创侵入性设备并发症的进阶监测能力。

要点

  • 单中心综合ICU、7例患者69次测量:光电容积描记胸贴设备与热稀释法心输出量高度相关(r=0.906)。
  • 平均心输出量:热稀释法7.3±2.0 L/min,光电容积描记法7.0±1.5 L/min。
  • Bland-Altman偏差0.3 L/min,95%一致性界限-1.6至2.2 L/min。
  • 百分比偏差2.4%,95%一致性界限-25.7%至30.5%。
  • 作者认为该设备与脉搏轮廓心输出量及热稀释法相比具有高准确性,可在多种临床场景提供无创进阶监测。

适用人群与场景

  • 综合ICU中按主治医师决定使用脉搏轮廓心输出量设备监测的患者。

处置流程与药物剂量

  • 平行使用光电容积描记胸贴设备与脉搏轮廓心输出量监测,医疗团队对无创设备获得的数值设盲。

研究设计与关键结果

  • 研究类型:观察性、前瞻性、比较临床试验;单中心综合ICU;最终分析7例患者共69次测量。
  • 干预与对照:光电容积描记胸贴设备与脉搏轮廓心输出量(热稀释法)平行监测,医疗团队对无创设备数值设盲。
  • 主要结果:平均心输出量热稀释法7.3±2.0 L/min vs 光电容积描记法7.0±1.5 L/min。
  • Bland-Altman分析:光电容积描记法偏差0.3 L/min,95%一致性界限-1.6至2.2 L/min;百分比偏差2.4%,95%一致性界限-25.7%至30.5%。
  • 相关性:热稀释法与光电容积描记法数值高度相关(r=0.906)。

证据类型、等级与局限

  • 证据类型:观察性、前瞻性、比较临床试验(简要报告)。
  • 局限:原文未报告局限性。

对床旁实践的改变

  • 作者认为,该胸贴设备与脉搏轮廓心输出量及热稀释法相比具有高准确性,可在多种临床场景提供进阶监测能力,且无侵入性设备的并发症。

提炼说明

  • 本条依据论文摘要整理;摘要未涉及「诊断标准与关键阈值」「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开

OBJECTIVES: Cardiac output (CO) measurements in the ICU are usually based on invasive techniques, which are technically complex and associated with clinical complications. This study aimed to compare CO measurements obtained from a noninvasive photoplethysmography-based device to a pulse contour cardiac output device in ICU patients.

DESIGN: Observational, prospective, comparative clinical trial.

SETTING: Single-center general ICU.

PATIENTS: Patients admitted to the general ICU monitored using a pulse contour cardiac output device as per the decision of the attending physician.

INTERVENTIONS: Parallel monitoring of CO using a photoplethysmography-based chest patch device and pulse contour cardiac output while the medical team was blinded to the values obtained by the noninvasive device.

MEASUREMENTS AND MAIN RESULTS: Seven patients (69 measurements) were included in the final analysis. Mean CO were 7.3 ± 2.0 L/m and 7.0 ± 1.5 L/m for thermodilution and photoplethysmography, respectively. Bland-Altman showed that the photoplethysmography has a bias of 0.3 L/m with -1.6 and 2.2 L/m 95% limit of agreement (LOA) and a bias of 2.4% with 95% LOA between -25.7% and 30.5% when calculating the percentage of difference from thermodilution. The values obtained by thermodilution and photoplethysmography were highly correlated (r = 0.906).

CONCLUSIONS: The tested chest patch device offers a high accuracy for CO compared to data obtained by the pulse contour cardiac output and the thermodilution method in ICU patients. Such devices could offer advanced monitoring capabilities in a variety of clinical settings, without the complications of invasive devices.

来源:Crit Care Explor,查看原文。版权归原作者所有。