使用安全铺巾进行手接触除颤:按压影响与替代电流通路分析
该研究在CPR人体模型上测量2 mil(0.002英寸)聚乙烯铺巾在30分钟持续胸外按压前后的泄漏电流,采用标准化IEC泄漏电流分析法和模拟电流经施救者手臂回流至患者的通路两种方法,每种通路各测10次。另对20例在埃默里大学医院因心房颤动(18/20)或心房扑动(2/20)接受择期心脏电复律的患者(年龄38-101岁,35%女性)均接受200 J电击进行分析。IEC法按压前峰值泄漏电流均值为0.70±0.02 mA,按压后为0.59±0.19 mA;评估经施救者手臂电流的10次测量中仅3次可检测到电流且幅度均低。所有测量均远低于IEC建议的最大值3.5 mA RMS和5.0 mA峰值。
要点
- 20例择期电复律患者(心房颤动18/20、心房扑动2/20)均接受200 J电击,年龄38-101岁,35%女性。
- IEC法测量峰值泄漏电流均值:按压前0.70±0.02 mA,30分钟持续按压后0.59±0.19 mA。
- 评估经施救者手臂电流的10次测量中仅3次可检测到电流,且每次幅度均低。
- 所有测量均远低于IEC建议的最大值3.5 mA RMS和5.0 mA峰值。
- 聚乙烯铺巾在长时间按压后仍可能有助于安全实施手接触除颤,经施救者手臂回路的电流幅度可能无显著意义。
适用人群与场景
- 在埃默里大学医院因心房颤动(18/20)或心房扑动(2/20)接受择期心脏电复律的患者,年龄38-101岁,35%女性,均接受200 J电击。
诊断标准与关键阈值
- IEC建议的最大泄漏电流限值:3.5 mA RMS和5.0 mA峰值。
研究设计与关键结果
- 研究在CPR人体模型上测量2 mil(0.002英寸)聚乙烯铺巾的泄漏电流,采用标准化IEC泄漏电流分析法和模拟电流经施救者手臂回流至患者的通路两种方法,每种通路各测10次,30分钟持续按压后重复测量。
- IEC法峰值泄漏电流均值:按压前0.70±0.02 mA,按压后0.59±0.19 mA。
- 评估经施救者手臂电流的10次测量中仅3次可检测到电流,且每次幅度均低。
- 所有测量均远低于IEC建议的最大值3.5 mA RMS和5.0 mA峰值。
- 20例择期电复律患者(心房颤动18/20、心房扑动2/20)均接受200 J电击,年龄38-101岁,35%女性。
证据类型、等级与局限
- 研究类型:在CPR人体模型上进行的体外测量研究,并纳入20例择期电复律患者分析。
- 局限:原文未报告局限性。
对床旁实践的改变
- 作者认为聚乙烯铺巾即使在长时间按压后仍可能有助于安全实施手接触除颤。
- 作者认为经施救者手臂回路的电流幅度可能无显著意义。
提炼说明
- 本条依据论文摘要整理;摘要未涉及「处置流程与药物剂量」「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开
BACKGROUND: Hands-on defibrillation (HOD) could theoretically improve the efficacy of cardiopulmonary resuscitation (CPR) though a few mechanisms. Polyethylene drapes could potentially facilitate safe HOD, but questions remain about the effects of CPR on polyethylene's conductance and the magnitude of current looping through rescuers' arms in contact with patients.
METHODS: This study measured the leakage current through 2 mil (0.002 in.) polyethylene through two different current pathways before and after 30 min of continuous compressions on a CPR mannequin. The two pathways analyzed were the standardized IEC (International Electrotechnical Commission) leakage current analysis and a setup analyzing a current pathway looping through a rescuer's arms and returning to the patient. First, ten measurements involving the two pathways were obtained on a single polyethylene drape. 30 min of continuous compressions were applied to the drape on a CPR mannequin after which the ten measurements were repeated.
RESULTS: Twenty patients undergoing elective cardioversion for atrial fibrillation (18/20) or atrial flutter (2/20) at Emory University Hospital underwent analysis all receiving 200 J shocks (age 38-101, 35% female). Through the IEC measurement method the peak leakage current mean was 0.70 +/- 0.02 mA before compressions and 0.59 +/- 0.19 mA after compressions. Only three of the ten measurements assessing current passing through a rescuer's arms had detectable current and each was of low magnitude. All measurements were well below the maximum IEC recommendations of 3.5 mA RMS and 5.0 mA peak.
CONCLUSIONS: Polyethylene may facilitate safe HOD even after long durations of compressions. Current looping through a rescuer's arms is likely of insignificant magnitude.