呼吸频率对呼吸振荡法测量值的影响:哮喘患者的随机交叉与观察性研究
研究在剑桥大学医院重症哮喘门诊招募59例哮喘患者,采用随机交叉设计在15、30、40次/分三种呼吸频率下进行强迫振荡技术检测,发现呼吸频率从15次/分增至40次/分时总气道阻力(Rrs at 5 Hz)显著降低8.6%(p<0.001),Delta R5–R19显著降低14.9%(p<0.001),5%参与者在较高频率下被重新分类为呼吸振荡法测量值在正常范围内;横断面分析纳入796例哮喘患者,自选呼吸频率中位数为16.55次/分(范围8.0–42.0),人口学资料与肺功能解释呼吸频率变异的23.5%。
要点
- 呼吸频率从15次/分增至40次/分时,总气道阻力(Rrs at 5 Hz)显著降低8.6%(p<0.001)。
- 较高呼吸频率下5%参与者被重新分类为呼吸振荡法测量值在正常范围内。
- 进行呼吸振荡法检测时,呼吸频率应控制在约15次/分以降低误分类风险。
核心结论
- 较高呼吸频率显著降低总Rrs,进行呼吸振荡法检测时呼吸频率应控制在约15次/分以降低误分类风险。
研究设计与样本
- 从剑桥大学医院重症哮喘门诊招募患者,采用随机交叉设计,参与者在15、30、40次/分三种呼吸频率下按随机顺序进行强迫振荡技术检测,并使用节拍器确保呼吸频率正确。
- 另利用哮喘患者横断面数据,采用多变量线性回归分析自选呼吸频率的决定因素。
- 随机交叉部分纳入59例患者,平均年龄54.1岁,58%为女性,中位FEV1/FVC z评分为–2.56(–3.26至–1.96)。
- 横断面分析提取796例哮喘患者数据。
主要发现
- 呼吸频率从15次/分增至40次/分时,总气道阻力(Rrs at 5 Hz)显著降低8.6%(p<0.001),且不由潮气量或流量变化介导。
- 呼吸频率从15次/分增至40次/分时,Delta R5–R19显著降低14.9%(p<0.001),电抗(Xrs at 5 Hz)增加10.9%(p=0.081)。
- 较高呼吸频率下,5%的研究参与者被重新分类为呼吸振荡法测量值在正常范围内。
- 横断面分析中,自选呼吸频率中位数(范围)为16.55次/分(8.0,42.0),人口学数据与肺功能解释呼吸频率变异的23.5%。
临床怎么用
- 进行呼吸振荡法检测时,呼吸频率应控制在约15次/分以降低误分类风险。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
Introduction The impact of breathing frequency on respiratory oscillometry measurements is unknown. We aimed to investigate the impact of different breathing frequencies in patients with asthma. Method We recruited patients from the severe asthma clinic at Cambridge University Hospitals. Using a randomised-crossover design, participants performed the forced oscillation technique at three different breathing frequencies (15, 30, 40 bpm) in a randomised order for each participant. A metronome was used to ensure the correct breathing frequency. Analysis of variance (ANOVA) and Friedman tests were utilised to assess the significance of any differences across the breathing frequencies. We then leveraged cross-sectional data from asthma patients who performed respiratory oscillometry and investigated the determinants of self-selected breathing frequency using multivariable linear regression. Results We recruited 59 patients. Mean age 54.1 years, 58% female, median forced expiratory volume in 1 s (FEV 1 )/FVC z-score –2.56 (–3.26 to –1.96). As breathing frequency increased from 15 bpm to 40 bpm, there was a significant reduction (–8.6%, p<0.001) in total airway resistance (Rrs at 5 Hz), which was not mediated by changes in tidal volume or flow rate. There was also a significant reduction (–14.9%, p<0.001) in Delta R5–R19, and an increase (+10.9%, p=0.081) in reactance (Xrs at 5 Hz). At higher breathing frequencies, 5% of study participants were reclassified as having respiratory oscillometry measurements within normal limits. Data from 796 asthma patients were extracted for the cross-sectional analysis. The median (range) self-selected breathing frequency was 16.55 bpm (8.0, 42.0). Demographic data and lung function explained 23.5% of the variation in breathing frequency. Conclusion Higher breathing frequencies significantly reduce total Rrs . When performing respiratory oscillometry, breathing frequency should be controlled at approximately 15 bpm to reduce risk of misclassification.
引用
- BMJ Open Respiratory Research:Impact of breathing frequency on respiratory oscillometry measurements: a randomised cross-over and observational study of asthma patients
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。