减少SARS-CoV-2气溶胶扩散的推荐无创通气回路设置可能导致呼吸机性能下降:基准比较研究
双肢回路联合口鼻面罩性能最佳,双肢回路联合头盔接口性能最差。
要点
- 回路设置显著影响无创通气性能,双肢回路加口鼻面罩性能最佳。
- 双肢回路加头盔接口性能最差,应避免优先选用。
- 选择减少气溶胶扩散的回路时需权衡呼吸机性能。
核心结论
- 呼吸机性能受回路设置显著影响,应优先使用双肢回路联合口鼻面罩。
研究设计与样本
- 台架测试研究,使用3D打印头模和人工肺评估8种回路设置在NIV和CPAP下的性能。
主要发现
- CPAP及加用细菌过滤器时也观察到类似差异。
- 双肢回路联合口鼻面罩性能最佳,双肢回路联合头盔接口性能最差。
临床怎么用
- 临床应优先选用双肢回路联合口鼻面罩进行无创通气,以兼顾气溶胶扩散减少与呼吸机性能。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: SARS-CoV-2 aerosolization during noninvasive positive-pressure ventilation may endanger health care professionals. Various circuit setups have been described to reduce virus aerosolization. However, these setups may alter ventilator performance. RESEARCH QUESTION: What are the consequences of the various suggested circuit setups on ventilator efficacy during CPAP and noninvasive ventilation (NIV)? STUDY DESIGN AND METHODS: Eight circuit setups were evaluated on a bench test model that consisted of a three-dimensional printed head and an artificial lung. Setups included a dual-limb circuit with an oronasal mask, a dual-limb circuit with a helmet interface, a single-limb circuit with a passive exhalation valve, three single-limb circuits with custom-made additional leaks, and two single-limb circuits with active exhalation valves. All setups were evaluated during NIV and CPAP. The following variables were recorded: the inspiratory flow preceding triggering of the ventilator, the inspiratory effort required to trigger the ventilator, the triggering delay, the maximal inspiratory pressure delivered by the ventilator, the tidal volume generated to the artificial lung, the total work of breathing, and the pressure-time product needed to trigger the ventilator. RESULTS: With NIV, the type of circuit setup had a significant impact on inspiratory flow preceding triggering of the ventilator (P < .0001), the inspiratory effort required to trigger the ventilator (P < .0001), the triggering delay (P < .0001), the maximal inspiratory pressure (P < .0001), the tidal volume (P = .0008), the work of breathing (P < .0001), and the pressure-time product needed to trigger the ventilator (P < .0001). Similar differences and consequences were seen with CPAP as well as with the addition of bacterial filters. Best performance was achieved with a dual-limb circuit with an oronasal mask. Worst performance was achieved with a dual-limb circuit with a helmet interface. INTERPRETATION: Ventilator performance is significantly impacted by the circuit setup. A dual-limb circuit with oronasal mask should be used preferentially.
引用
- Patout M, Fresnel E, Lujan M, et al. Recommended Approaches to Minimize Aerosol Dispersion of SARS-CoV-2 During Noninvasive Ventilatory Support Can Cause Ventilator Performance Deterioration: A Benchmark Comparative Study. Chest. 2021. PMID: 33667491. doi:10.1016/j.chest.2021.02.047
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。