无创通气期间新型沟通设备的随机对照试验

Randomized Controlled Trial of a Novel Communication Device Assessed During Noninvasive Ventilation Therapy.
PubMed 文献 · Chest · 2020-10-01

一项单中心随机对照试验纳入36例接受CPAP治疗的OSA门诊患者,评估口鼻面罩对沟通清晰度的影响及沟通设备(SPEAX)的改善作用。结果显示,与正常对话相比,CPAP治疗时单词和句子清晰度分别相对下降52%和57%;使用设备后,干预组单词和句子清晰度分别较对照组相对提高75%和126%(P < .001),患者和伴侣感知的舒适度也显著改善。

要点

  • CPAP治疗时沟通清晰度显著下降,单词和句子清晰度分别相对降低52%和57%。
  • 使用沟通设备可显著提高单词和句子清晰度(分别相对提高75%和126%,P < .001)。
  • 该设备还改善了患者及伴侣感知的舒适度(P = .042和P = .0074)。

核心结论

  • 使用该PAP沟通设备可显著改善沟通清晰度和舒适度。
  • 这是首批量化PAP治疗期间沟通障碍的研究之一。

研究设计与样本

  • 单中心随机对照试验,纳入36例接受CPAP治疗的OSA门诊患者。
  • 每对患者-伴侣进行三次评估:基线正常对话、面罩基线(CPAP下对话)、随机使用功能设备或假设备。
  • CPAP压力为10 cm H2O,沟通设备为SPEAX(Ataia Medical)。
  • 通过阅读单词和句子由伴侣记录正确率(%words correct)评估沟通清晰度。

主要发现

  • 与正常对话相比,CPAP下单词清晰度相对下降52%(87% vs 41%),句子清晰度相对下降57%(94% vs 40%)。
  • 干预组单词清晰度相对提高75%(35% vs 61%;P < .001),句子清晰度相对提高126%(33% vs 76%;P < .001),均高于对照组。
  • 设备使患者感知的PAP舒适度相对提高233%(15% vs 50%,P = .042),伴侣感知舒适度相对提高245%(20% vs 69%,P = .0074)。

临床怎么用

  • 对于接受无创通气的患者,可考虑使用沟通设备以改善沟通清晰度和舒适度。

局限性

  • 单中心研究,样本量较小(36例门诊OSA患者),结果外推需谨慎。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Noninvasive ventilation (NIV), a form of positive airway pressure (PAP) therapy, is the standard of care for various forms of acute respiratory failure (ARF). Communication impairment is a side effect of NIV, impedes patient care, contributes to distress and intolerance, and potentially increases intubation rates. This study aimed to evaluate communication impairment during CPAP therapy and demonstrate communication device improvement with a standardized protocol. RESEARCH QUESTION: How does an oronasal mask affect communication intelligibility? How does use of an NIV communication device change this communication intelligibility? STUDY DESIGN AND METHODS: A single-center randomized controlled trial (36 outpatients with OSA on CPAP therapy) assessed exposure to CPAP 10 cm H2O and PAP communication devices (SPEAX, Ataia Medical). Communication impairment was evaluated by reading selected words and sentences for partners to record and were tabulated as %words correct. Each outpatient-partner pair performed three assessments: (1) baseline (conversing normally), (2) mask baseline (conversing with PAP), and (3) randomized to functioning device (conversing with PAP and device) or sham device. After each stage, both outpatients and partners completed Likert surveys regarding perceived intelligibility and comfort. RESULTS: While conversing with PAP, word and sentence intelligibility decreased relatively by 52% (87% vs 41%) and relatively by 57% (94% vs 40%), respectively, compared with normal conversation. Word and sentence intelligibility in the intervention arm increased relatively by 75% (35% vs 61%; P < .001) and by 126% (33% vs 76%; P < .001) higher than the control arm, respectively. The device improved outpatient-perceived PAP comfort relatively by 233% (15% vs 50%, P = .042) and partner-perceived comfort by relatively 245% (20% vs 69%, P = .0074). INTERPRETATION: Use of this PAP communication device significantly improves both intelligibility and comfort. This is one of the first studies quantifying communication impairment during PAP delivery. TRIAL REGISTRY: ClinicalTrials.gov; No.: NCT03795753; URL: www.clinicaltrials.gov.