文献速递睡眠呼吸障碍

AVAPS-AE与ST模式在肥胖低通气综合征中的随机对照试验

AVAPS-AE versus ST mode: A randomized controlled trial in patients with obesity hypoventilation syndrome.
PubMed 文献 · Respirology · 2020-02-13

这项前瞻性多中心随机对照试验比较了AVAPS-AE与ST模式通气2个月对稳定期肥胖低通气综合征(OHS)患者客观睡眠质量的影响。结果显示两组在睡眠质量指标上无显著差异,PaCO2改善程度相似,AVAPS-AE组NIV设置时间更短。

要点

  • AVAPS-AE与ST通气2个月对OHS患者睡眠质量和血气交换影响相似。
  • 两组PaCO2平均降低均为-0.87 kPa,组间无显著差异。
  • AVAPS-AE组NIV设置时间更短。

核心结论

  • AVAPS-AE与ST通气2个月对稳定期OHS患者的睡眠质量和血气交换具有相似影响。

研究设计与样本

  • 前瞻性多中心随机对照试验,纳入69例OHS患者,60例成功设置NIV,最终26例ST组和30例AVAPS-AE组有基线和随访PSG数据。

主要发现

  • 基线PaCO2在ST组为6.94±0.71 kPa,AVAPS-AE组为6.61±0.71 kPa(P=0.032)。
  • 两组客观睡眠质量指标无显著组间差异。
  • PaCO2平均降低ST组-0.87 kPa(95%CI:-1.12至-0.46),AVAPS-AE组-0.87 kPa(95%CI:-1.14至-0.50),P=0.984。
  • 两组平均NIV使用时间均为每晚6.2小时(P=0.93)。
  • AVAPS-AE组NIV设置时间更短(P=0.012)。

关键数值与推荐

  • 纳入标准:日间PaCO2>6 kPa,BMI≥30 kg/m2,临床稳定超过2周,且未接受过家庭NIV。

临床怎么用

  • 对于稳定期OHS患者,AVAPS-AE与ST模式在睡眠质量和血气改善方面效果相当,AVAPS-AE可能简化NIV设置。

局限性

  • 最终分析样本量较小(ST组26例,AVAPS-AE组30例)。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND AND OBJECTIVE: Average volume-assured pressure support-automated expiratory positive airway pressure (AVAPS-AE) combines an automated positive expiratory pressure to maintain upper airway patency to an automated pressure support with a targeted tidal volume. The aim of this study was to compare the effects of 2-month AVAPS-AE ventilation versus pressure support (ST) ventilation on objective sleep quality in stable patients with OHS. Secondary outcomes included arterial blood gases, health-related quality of life, daytime sleepiness, subjective sleep quality and compliance to NIV. METHODS: This is a prospective multicentric randomized controlled trial. Consecutive OHS patients included had daytime Pa CO2 > 6 kPa, BMI ≥ 30 kg/m2 , clinical stability for more than 2 weeks and were naive from home NIV. PSG were analysed centrally by two independent experts. Primary endpoint was sleep quality improvement at 2 months. RESULTS: Among 69 trial patients, 60 patients had successful NIV setup. Baseline and follow-up PSG were available for 26 patients randomized in the ST group and 30 in the AVAPS-AE group. At baseline, Pa CO2 was 6.94 ± 0.71 kPa in the ST group and 6.61 ± 0.71 in the AVAPS-AE group (P = 0.032). No significant between-group difference was observed for objective sleep quality indices. Improvement in Pa CO2 was similar between groups with a mean reduction of -0.87 kPa (95% CI: -1.12 to -0.46) in the ST group versus -0.87 kPa (95% CI: -1.14 to -0.50) in the AVAPS-AE group (P = 0.984). Mean NIV use was 6.2 h per night in both groups (P = 0.93). NIV setup duration was shorter in the AVAPS-AE group (P = 0.012). CONCLUSION: AVAPS-AE and ST ventilation for 2 months had similar impact on sleep quality and gas exchange.

来源:Respirology,查看原文。版权归原作者所有。

引用

  1. Patout M, Gagnadoux F, Rabec C, et al. AVAPS-AE versus ST mode: A randomized controlled trial in patients with obesity hypoventilation syndrome. Respirology. 2020. PMID: 32052923. doi:10.1111/resp.13784

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。