肺癌筛查CT:按性别估算有效辐射剂量的转换因子
该回顾性研究纳入2016至2017年某大型医疗系统的1890例肺癌筛查患者,用蒙特卡洛模拟计算每例有效剂量并除以剂量长度乘积得出k因子。全组平均k因子为0.0179 mSv·mGy⁻¹·cm⁻¹,较既往胸部CT标准值0.014高22%;女性为0.0213、男性为0.0149,女性比男性高43%。建议肺癌筛查采用性别特异k因子估算有效剂量。
要点
- 肺癌筛查低剂量CT的k因子全组均值为0.0179 mSv·mGy⁻¹·cm⁻¹,高于胸部CT标准值0.014。
- 女性k因子0.0213 mSv·mGy⁻¹·cm⁻¹,男性0.0149 mSv·mGy⁻¹·cm⁻¹,女性比男性高43%。
- 估算肺癌筛查有效辐射剂量应采用性别及筛查特异k因子,而非通用胸部CT值。
核心结论
- 肺癌筛查低剂量CT的k因子高于既往胸部CT通用值,且女性比男性高43%,应采用性别及筛查特异k因子估算有效剂量。
研究设计与样本
- 回顾性研究,纳入2016至2017年某大型医疗系统连续肺癌筛查患者共1890例。
- 通过患者特异蒙特卡洛模拟(Cristy体模)计算有效剂量,除以研究剂量长度乘积得出k因子。
- 用单样本Student t检验将k因子与胸部CT标准值0.014 mSv·mGy⁻¹·cm⁻¹比较,并行双变量和多变量分析。
主要发现
- 全组平均k因子为0.0179 mSv·mGy⁻¹·cm⁻¹,比胸部CT标准值0.014高22%(P < .001)。
- 多变量模型中女性平均k因子为0.0213 mSv·mGy⁻¹·cm⁻¹,男性为0.0149 mSv·mGy⁻¹·cm⁻¹,女性比男性高43%(P < .001)。
关键数值与推荐
- 胸部CT通用k因子标准值为0.014 mSv·mGy⁻¹·cm⁻¹。
- 肺癌筛查全组k因子均值0.0179 mSv·mGy⁻¹·cm⁻¹,女性0.0213 mSv·mGy⁻¹·cm⁻¹,男性0.0149 mSv·mGy⁻¹·cm⁻¹。
临床怎么用
- 肺癌筛查项目估算有效辐射剂量时,应使用性别特异及肺癌筛查特异的k因子,避免沿用通用胸部CT的0.014值。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Effective dose (ED) is used to understand radiation-related cancer risk of CT scans. Currently, ED for low-dose CT (LDCT) lung cancer screening (LCS) is estimated by multiplying the CT scan-reported dose-length product (DLP) by a DLP-to-ED conversion factor (k-factor) for general chest CT imaging, which does not account for sex. The purpose of this study was to calculate sex-specific k-factors for LDCT LCS. METHODS: This retrospective study evaluated consecutive LCS patients across a large health system from 2016 to 2017. Patient and CT scan-related data were obtained from the radiology information system, the picture archiving and communication system, and a radiation dose index-monitoring system. Each patient's ED was determined by patient-specific Monte-Carlo simulation using Cristy phantoms and divided by study DLP to determine the k-factor. The k-factors were compared vs the standard of 0.014 mSv·mGy⁻1·cm⁻1 for a chest CT scan by using a one-sample Student t test. Bivariate and multivariable analyses were performed for k-factors based on patient and CT scan factors. RESULTS: A total of 1,890 patients were included in the study. The mean k-factor for all patients was 0.0179 mSv·mGy⁻1·cm⁻1, which was 22% greater than the standard value of 0.014 mSv·mGy⁻1·cm⁻1 for a chest CT scan previously applied to LDCT imaging (P < .001). The mean k-factor in women (0.0213 mSv·mGy⁻1·cm⁻1) was 43% greater than in men (0.0149 mSv·mGy⁻1·cm⁻1) in the multivariable model (P < .001). CONCLUSIONS: The overall k-factor for LCS is higher than the previously used value for chest CT imaging; when stratified according to sex, it was 43% greater in women than in men. Sex- and LCS-specific k-factors should be used to estimate effective radiation dose in LCS programs.