文献速递呼吸危重症与机械通气呼吸道感染

高流量鼻导管与不同接口CPAP治疗时呼出气扩散范围比较

Exhaled air dispersion during high-flow nasal cannula therapy versus CPAP via different masks.
PubMed 文献 · Eur Respir J · 2019-04-11

本研究在人患者模拟器上比较了高流量鼻导管(HFNC)与经鼻枕或口鼻面罩持续气道正压(CPAP)治疗时的呼出气扩散距离。结果显示,在接口贴合良好时,HFNC和CPAP的呼出气扩散均有限;但HFNC管路松动时侧向泄漏可达620 mm。

要点

  • CPAP经鼻枕时呼出气扩散随压力升高而增加,口鼻面罩泄漏可忽略。
  • HFNC流量从10升至60 L·min-1时呼出气扩散距离增加。
  • HFNC接口管松动时侧向空气泄漏可达620 mm。

核心结论

  • 在接口贴合良好的前提下,HFNC和不同接口CPAP的呼出气扩散均有限。

研究设计与样本

  • 在人患者模拟器(HPS)上于16次换气·h-1的隔离房间内比较HFNC与CPAP的呼出气扩散。
  • HPS设定为模拟不同严重程度的肺损伤。

主要发现

  • 正常肺条件下,经Respironics鼻枕CPAP从5升至20 cmH2O时,矢状面呼出气扩散距离从186±34增至264±27 mm。
  • 经ResMed鼻枕CPAP从5升至20 cmH2O时,扩散距离从207±11增至332±34 mm。
  • 口鼻面罩泄漏可忽略。
  • HFNC从10升至60 L·min-1时,呼出气距离从65±15增至172±33 mm。
  • HFNC及接口管松动时,侧向空气泄漏可达620 mm。

关键数值与推荐

  • 显著暴露定义为归一化烟雾浓度≥20%。
  • CPAP经鼻枕以5-20 cmH2O输送,HFNC以10-60 L·min-1输送,后者加温湿化至37°C。

临床怎么用

  • 临床使用HFNC或CPAP时需确保接口贴合良好以减少呼出气扩散。
  • 应避免HFNC管路松动,防止侧向空气泄漏增加暴露风险。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: High-flow nasal cannula (HFNC) is an emerging therapy for respiratory failure but the extent of exhaled air dispersion during treatment is unknown. We examined exhaled air dispersion during HFNC therapy versus continuous positive airway pressure (CPAP) on a human patient simulator (HPS) in an isolation room with 16 air changes·h-1. METHODS: The HPS was programmed to represent different severity of lung injury. CPAP was delivered at 5-20 cmH2O via nasal pillows (Respironics Nuance Pro Gel or ResMed Swift FX) or an oronasal mask (ResMed Quattro Air). HFNC, humidified to 37°C, was delivered at 10-60 L·min-1 to the HPS. Exhaled airflow was marked with intrapulmonary smoke for visualisation and revealed by laser light-sheet. Normalised exhaled air concentration was estimated from the light scattered by the smoke particles. Significant exposure was defined when there was ≥20% normalised smoke concentration. RESULTS: In the normal lung condition, mean±sd exhaled air dispersion, along the sagittal plane, increased from 186±34 to 264±27 mm and from 207±11 to 332±34 mm when CPAP was increased from 5 to 20 cmH2O via Respironics and ResMed nasal pillows, respectively. Leakage from the oronasal mask was negligible. Mean±sd exhaled air distances increased from 65±15 to 172±33 mm when HFNC was increased from 10 to 60 L·min-1. Air leakage to 620 mm occurred laterally when HFNC and the interface tube became loose. CONCLUSION: Exhaled air dispersion during HFNC and CPAP via different interfaces is limited provided there is good mask interface fitting.