便携式制氧机与氧气钢瓶在间质性肺疾病患者步行中的随机交叉试验
本研究纳入20例运动后血氧饱和度低于90%的间质性肺疾病(ILD)患者,采用随机交叉设计比较两种便携式制氧机(Inogen One G2和EverGo)与压缩氧气钢瓶在6分钟步行试验中的表现。结果显示三种设备在最低SpO2和6分钟步行距离上均无显著差异,提示两种便携式制氧机在步行中的性能与氧气钢瓶相当。
要点
- 两种便携式制氧机在ILD患者步行中的最低SpO2与氧气钢瓶无显著差异。
- 三种设备的6分钟步行距离无显著差异。
- 便携式制氧机可作为ILD患者运动时氧气钢瓶的替代选择。
核心结论
- Inogen One G2和EverGo便携式制氧机在间质性肺疾病运动后低氧患者步行中的表现与压缩氧气钢瓶相当。
研究设计与样本
- 随机交叉试验,纳入20例不同病因的间质性肺疾病患者,均在6分钟步行试验中室内空气下出现SpO2低于90%的运动性低氧。
主要发现
- 试验1中Inogen One G2制氧机与氧气钢瓶的最低SpO2分别为82.3±3.5%和80.3±2.2%,P=0.14。
- 试验2中EverGo制氧机与氧气钢瓶的最低SpO2分别为85.7±7.7%和86.1±6.1%,P=0.79。
- 三种设备的平均6分钟步行距离无显著差异。
临床怎么用
- 对于间质性肺疾病伴运动性低氧患者,可考虑使用便携式制氧机替代压缩氧气钢瓶进行步行时的氧疗。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND AND OBJECTIVE: Ambulatory oxygen therapy is often provided to patients with interstitial lung disease (ILD). Lightweight portable oxygen concentrators (POCs) provide an alternative to traditional portable systems such as compressed oxygen cylinders; however, their efficacy in patients with ILD has not been assessed. This study aimed to evaluate the clinical performance of three ambulatory oxygen systems (two different POCs and a compressed oxygen cylinder) during 6-min walk tests (6MWTs) in patients with ILD and exertional desaturation. METHODS: A total of 20 participants with ILD of varying aetiologies who demonstrated exertional desaturation to <90% on room air during 6MWT were recruited. Each participant performed two 6MWTs while breathing room air. On a subsequent day, two further 6MWTs were performed, in random order: one breathing oxygen via a POC (either the Inogen One G2 POC or the EverGo POC at the setting of 6) and one with a compressed oxygen cylinder (at 5 L/min). RESULTS: There were no significant differences in nadir oxygen saturation (SpO2 ) during 6MWTs using different portable oxygen devices (Trial 1: mean SpO2 for Inogen One G2 POC: 82.3 ± 3.5% vs oxygen cylinder: 80.3 ± 2.2%, P = 0.14; Trial 2: mean SpO2 for EverGo POC: 85.7 ± 7.7% vs oxygen cylinder: 86.1 ± 6.1%, P = 0.79). The mean 6-min walk distances were not significantly different among the three devices. CONCLUSION: The performance of the Inogen One G2 POC and the EverGo POC had comparable performance with that of the compressed oxygen cylinder during walking in patients with ILD and exertional desaturation.
引用
- Khor YH, McDonald CF, Hazard A, et al. Portable oxygen concentrators versus oxygen cylinder during walking in interstitial lung disease: A randomized crossover trial. Respirology. 2017. PMID: 28544460. doi:10.1111/resp.13083