文献速递睡眠呼吸障碍

新型经皮舌下神经刺激对OSA气流及气道塌陷性的影响

Effect of Novel Hypoglossal Stimulation on Airflow and Airway Collapsibility in OSA.
PubMed 文献 · Chest · 2025-09-06

研究纳入14例中重度OSA患者,经超声引导经皮植入线性多对电极阵列行舌下神经刺激(HNS)。HNS使13例患者气流增加,峰值吸气流速恢复至治疗性CPAP水平,并改善气道塌陷性(Pcrit从0.1降至-7.2 cm H2O)。

要点

  • 经皮超声引导HNS可急性增加OSA患者气流至治疗性CPAP等效水平。
  • HNS使峰值吸气流速从0.10 L/s恢复至0.41 L/s,停止后回落至0.14 L/s。
  • HNS改善气道塌陷性,主动Pcrit从0.1降至-7.2 cm H2O。

核心结论

  • 经皮超声引导的急性HNS可增加气流至治疗性CPAP等效水平并改善OSA患者气道塌陷性。

研究设计与样本

  • 前瞻性研究,纳入14例接受丙泊酚药物诱导睡眠内镜的中重度OSA患者(平均AHI 30 events/h)。
  • 经超声识别舌下神经并引导经皮植入线性多对电极阵列,测试HNS振幅0.5-5 mA及电极组合。

主要发现

  • HNS使13例参与者气流增加。
  • 治疗性CPAP下平均峰值吸气流速为0.43 L/s;CPAP短暂降低时第1-3次呼吸降至0.10 L/s,HNS期间第4-6次呼吸升至0.41 L/s,停止后第7-9次呼吸降至0.14 L/s(P < .001)。
  • HNS时主动Pcrit为-7.2 cm H2O,无HNS时为0.1 cm H2O(P < .01)。

关键数值与推荐

  • HNS振幅测试范围0.5-5 mA。
  • CPAP短暂降低诱导气流受限或阻塞≥9次呼吸/努力,HNS应用于第4至6次呼吸。

临床怎么用

  • 经皮超声引导HNS可能作为微创疗法用于中重度OSA,急性期可恢复气流并改善气道塌陷性。

局限性

  • 样本量小(14例),为急性研究,长期疗效与安全性未评估。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Hypoglossal nerve stimulation (HNS) to treat OSA currently requires placement of a cuff or "saddle" electrode around or adjacent to the hypoglossal nerve or nerves. Limitations for this therapy include cost, invasiveness, and variable efficacy. RESEARCH QUESTION: Can HNS applied via percutaneous implantation of a linear, multipair electrode array restore airflow to airway narrowing, obstruction, or both and improve airway collapsibility in people with OSA? STUDY DESIGN AND METHODS: Participants with OSA undergoing drug-induced sleep endoscopy with propofol were instrumented with an epiglottic pressure catheter, nasal mask, and pneumotachograph. Ultrasound was used to identify the hypoglossal nerve and to guide percutaneous electrode array placement. Transient CPAP reductions induced airflow limitation or obstruction for ≥ 9 breaths/efforts with HNS applied during breaths 4 through 6. A range of HNS amplitudes (0.5-5 mA) and electrode array combinations were tested to determine optimal airflow responses. Time-permitting, active critical closing pressure (Pcrit) also was quantified with and without HNS. RESULTS: Fourteen people with moderate to severe OSA (mean [SD] apnea-hypopnea index, 30 [16] events/h) were studied. HNS increased airflow in 13 participants. Mean (SD) peak inspiratory flow on therapeutic CPAP was 0.43 (0.09) L/s. During transient CPAP reductions, mean (SD) peak inspiratory flow reduced to 0.10 (0.07) L/s for breaths 1 through 3, increased to therapeutic CPAP levels with HNS for breaths 4 through 6 (0.41 (0.18) L/s), and returned to 0.14 (0.14) L/s for breaths 7 through 9 after HNS (P < .001). Mean (SD) active Pcrit with HNS vs without HNS was -7.2 (1.6) cm H2O vs 0.1 (2.4) cm H2O, respectively (P < .01). INTERPRETATION: Our results show that acute HNS via a minimally invasive, percutaneous, ultrasound-guided approach increases airflow to levels equivalent to therapeutic CPAP and improves airway collapsibility in people with OSA.

来源:Chest,查看原文。版权归原作者所有。

引用

  1. Osman AM, Carney AS, Ooi EH, et al. Effect of Novel Hypoglossal Stimulation on Airflow and Airway Collapsibility in OSA. Chest. 2025. PMID: 40921399. doi:10.1016/j.chest.2025.08.020

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。