机器人支气管镜针基共聚焦激光显微内镜诊断外周肺结节
本研究评估机器人支气管镜下针基共聚焦激光显微内镜(nCLE)实时引导诊断小外周肺结节的可行性、安全性及基于nCLE的针调整。20例疑似外周肺癌患者中,无并发症发生,19/20获得高质量nCLE视频,9例(45%)因nCLE显示针位置不佳而调整,调整后19/20确认针在结节内,总体诊断率80%(16/20)。
要点
- 机器人支气管镜nCLE成像可行且安全,无并发症发生。
- nCLE实时反馈使45%(9/20)患者调整活检针位置。
核心结论
- 机器人支气管镜nCLE成像在小外周肺结节中可行且安全,能提供实时反馈优化采样位置并可能提高诊断率。
研究设计与样本
- 前瞻性研究纳入20例疑似外周肺癌患者,行透视和径向EBUS辅助机器人支气管镜,结节识别后行nCLE成像。
- 根据nCLE恶性及气道/肺实质标准确定最佳采样位置,若见气道则调整活检针,nCLE确认针在结节内后于同位置活检。
主要发现
- 20例患者中最终诊断肺癌17例,肺结节中位大小14.5 mm(范围8-28 mm),无并发症发生。
- 19/20患者获得高质量nCLE视频,9例(45%)实时nCLE显示针位置不佳并调整。
- 调整后nCLE确认针在结节内19/20例,快速现场评估(ROSE)显示 representative 材料9/20例(45%),总体诊断率80%(16/20)。
- 3例nCLE恶性但病理不足者中,2例随访诊断为恶性。
关键数值与推荐
- 采用nCLE恶性及气道/肺实质标准识别最佳采样位置,若见气道则调整活检针。
- nCLE确认针在结节内后于同位置行针吸和活检。
临床怎么用
- 在机器人支气管镜诊断小外周肺结节时,可结合nCLE实时成像反馈调整活检针位置以优化采样。
局限性
- 样本量小(20例),需更大研究验证诊断率提升程度。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND AND OBJECTIVE: Robotic bronchoscopy has demonstrated high navigational success in small peripheral lung nodules but the diagnostic yield is discrepantly lower. Needle based confocal laser endomicroscopy (nCLE) enables real-time microscopic imaging at the needle tip. We aim to assess feasibility, safety and needle repositioning based on real-time nCLE-guidance during robotic bronchoscopy in small peripheral lung nodules. METHODS: Patients with suspected peripheral lung cancer underwent fluoroscopy and radial EBUS assisted robotic bronchoscopy. After radial EBUS nodule identification, nCLE-imaging of the target area was performed. nCLE-malignancy and airway/lung parenchyma criteria were used to identify the optimal sampling location. In case airway was visualized, repositioning of the biopsy needle was performed. After nCLE tool-in-nodule confirmation, needle passes and biopsies were performed at the same location. MEASUREMENTS AND MAIN RESULTS: Twenty patients were included (final diagnosis n = 17 (lung) cancer) with a median lung nodule size of 14.5 mm (range 8-28 mm). No complications occurred. In 19/20 patients, good quality nCLE-videos were obtained. In 9 patients (45%), real-time nCLE-imaging revealed inadequate positioning of the needle and repositioning was performed. After repositioning, nCLE-imaging provided tool-in-nodule-confirmation in 19/20 patients. Subsequent ROSE demonstrated representative material in 9/20 patients (45%) and overall diagnostic yield was 80% (16/20). Of the three patients with malignant nCLE-imaging but inadequate pathology, two were diagnosed with malignancy during follow-up. CONCLUSION: Robotic bronchoscopic nCLE-imaging is feasible and safe. nCLE-imaging in small, difficult-to-access lung nodules provided additional real-time feedback on the correct needle positioning with the potential to optimize the sampling location and diagnostic yield.
引用
- Manley CJ, Kramer T, Kumar R, et al. Robotic bronchoscopic needle-based confocal laser endomicroscopy to diagnose peripheral lung nodules. Respirology. 2022. PMID: 36535801. doi:10.1111/resp.14438
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。