文献速递呼吸危重症与机械通气

密闭式延长管对人工气道吸痰期间预防气道漏气的作用

[Effect of closed extension tube on preventing airway leakage during artificial airway clearance].
PubMed 文献 · Zhonghua Jie He He Hu Xi Za Zhi · 2022-01-01

该研究采用人工模肺连接呼吸机进行机械通气,比较连接开放式与密闭式延长管在吸痰作业期间气道压力和潮气量的变化。结果显示密闭式延长管可显著减少气道压力与潮气量下降,降低气道漏气,值得临床推荐。

要点

  • 密闭式延长管在吸痰时气道压力下降最大仅(0.83±0.27)cmH2O,潮气量下降最大(0.008±0.006)L。
  • 开放式延长管吸痰口开放时,吸气压力下降可达(14.82±0.51)cmH2O,潮气量下降可达(0.164±0.021)L。
  • 呼吸机设置压力越高,开放式延长管漏气越明显,密闭式延长管可稳定通气回路。

核心结论

  • 机械通气回路中连接密闭式延长管可降低人工气道吸痰作业期间的气道漏气,值得临床推荐使用。

研究设计与样本

  • 采用人工模肺连接呼吸机进行机械通气,呼吸回路分别连接开放式延长管或密闭式延长管。
  • 呼吸机压力参数设置为16/5、20/6、24/7、28/8、32/9、36/10 cmH2O六组递增压力,观察吸痰作业期间气道压力和潮气量变化。

主要发现

  • 相同呼吸机参数下,连接开放式延长管的气道压力均低于连接密闭式延长管(均P<0.05),且随设置压力增加差值逐渐增大,最高吸气压力36 cmH2O时差值达1.69 cmH2O。
  • 开放式延长管吸痰口开放时,气道压力和潮气量均显著下降(均P<0.05),且随呼吸机设置压力增加下降幅度增大;设置参数36/10 cmH2O时,吸气压力下降(14.82±0.51)cmH2O,潮气量下降最大(0.164±0.021)L。
  • 从开放式延长管插拔口插入吸痰管时,气道压力和潮气量较开放状态增加但仍低于关闭状态(均P<0.05),与关闭状态比较压力下降最大(3.53±0.46)cmH2O,潮气量下降最大(0.101±0.011)L。
  • 从密闭式延长管插拔口插入吸痰管期间,气道压力下降最大(0.83±0.27)cmH2O,潮气量下降最大(0.008±0.006)L,下降值均明显低于相同压力参数下开放式延长管(均P<0.05)。

关键数值与推荐

  • 呼吸机压力参数设置为16/5、20/6、24/7、28/8、32/9、36/10 cmH2O(1 cmH2O=0.098 kPa)。
  • 连接开放式延长管时,最高吸气压力36 cmH2O下与密闭式延长管的气道压力差值达1.69 cmH2O。
  • 开放式延长管吸痰口开放时,36/10 cmH2O设置下吸气压力下降(14.82±0.51)cmH2O,潮气量下降最大(0.164±0.021)L。
  • 密闭式延长管插管吸痰时,气道压力下降最大(0.83±0.27)cmH2O,潮气量下降最大(0.008±0.006)L。

临床怎么用

  • 在机械通气患者进行人工气道吸痰时,建议使用密闭式延长管以减少气道漏气和潮气量损失。
  • 对于呼吸机设置压力较高的患者,开放式延长管漏气更明显,更应考虑密闭式延长管。

局限性

  • 研究采用人工模肺进行体外测试,非临床患者,结果需临床验证。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

Objective: To explore the role of closed extension tube in preventing airway leakage during artificial airway clearance. Methods: The test lung was connected with a ventilator for mechanical ventilation. The pressure parameters were set as 16/5, 20/6, 24/7, 28/8, 32/9 and 36/10 cmH2O(1 cmH2O=0.098 kPa), respectively. The circuit was connected with an open extension tube or a closed extension tube. The ventilator was set with different pressure parameters to observe the changes of airway pressure and tidal volume during airway clearance. Results: (1) The pressure parameters were set as 16/5, 20/6, 24/7, 28/8, 32/9 and 36/10 cmH2O, and the airway pressures (in cmH2O) of circuit connected with open extension tube were (15.94±0.27)/(4.81±0.04), (20.09±0.23)/(6.05±0.16), (23.89±0.41)/(6.94±0.06), (27.90±0.22)/(7.71±0.18), (31.92±0.13)/(8.74±0.12)and(35.65±0.31)/(9.72±0.07), respectively.Under the same ventilator pressure parameters, the airway pressures (in cmH2O) of circuit connected with close extension tube were (16.36±0.06)/(4.85±0.04), (20.54±0.26)/(6.44±0.12), (24.36±0.24)/(7.01±0.33), (28.69±0.25)/(8.07±0.08), (32.97±0.33)/(8.93±0.09), (37.34±0.29)/(9.75±0.08), respectively. The airway pressure of circuit connected with open extension tube was lower than that connected with closed extension tube(P<0.05);with the increase of the pressure setting of the ventilator, the difference of the airway pressure between the two extended tubes gradually increased. When the maximum inspiratory pressure of the ventilator was set 36 cmH2O, the difference reached 1.69 cmH2O. (2) The airway pressures (in cmH2O) dropped from (15.94±0.27)/(4.81±0.04), (20.09±0.23)/(6.05±0.16), (23.89±0.41)/(6.94±0.06), (27.90±0.22)/(7.71±0.18), (31.92±0.13)/(8.74±0.12), (35.65±0.31)/(9.72±0.07) to (13.42±0.4)/(3.15±0.14), (16.81±0.6)/(4.30±0.14), (20.22±0.5)/(5.48±0.45), (23.73±1.4)/(6.25±0.22), (24.78±0.7)/(7.13±0.21), (20.83±0.4)/(6.61±0.19)when the suction port of the open extension tube was opened (P<0.05);and the tidal volume (in L) also decreased from 0.328±0.004, 0.580±0.012, 0.621±0.003, 0.626±0.003, 0.615±0.003, 0.603±0.002 to 0.272±0.008, 0.416±0.051, 0.487±0.047, 0.396±0.116, 0.507±0.022, 0.508±0.079, respectively (P<0.05). The decrease of airway pressure and tidal volume gradually increased with the increase of ventilator setting pressure. When the ventilator setting parameter was 36/10 cmH2O, the decrease of airway inspiratory pressure was (14.82±0.51) cmH2O and the maximum reduction of tidal volume was (0.164±0.021)L. (3)The airway pressure (in cmH2O) was increased to(15.70±0.23)/(4.80±0.33), (19.01±0.81)/(5.71±0.34), (22.27±0.62)/(6.85±0.44), (25.35±2.09)/(7.94±0.16), (28.38±0.46)/(8.96±0.23), (33.34±0.71)/(9.71±0.25) when the suction tube was inserted from the suction port of the open extension tube in the open state, and the tidal volume (in L) was increased to 0.340±0.016, 0.563±0.020, 0.571±0.030, 0.556±0.026, 0.514±0.021, 0.512±0.031 as well.The airway pressure and tidal volume of the ventilation circuit were higher than those in the open state, but still lower than those in the closed state. Compared with the closed state of the suction port, the maximum pressure drop and tidal volume decrease were (3.53±0.46) cmH2O and (0.101±0.011) L, respectively. (4) The pressure of the ventilator was set between 16/5 cmH2O to 36/10 cmH2O. The airway pressure (in cmH2O) was decreased from (16.26±0.04)/(4.85±0.04), (20.74±0.15)/(6.42±0.11), (25.09±0.31)/(7.10±0.13), (29.38±0.24)/(8.17±0.09), (33.80±0.16)/(9.02±0.17), (37.89±0.19)/(9.83±0.07) to(16.36±0.06)/(4.85±0.04), (20.54±0.26)/(6.44±0.12), (24.36±0.24)/(7.01±0.33), (28.69±0.25)/(8.07±0.08), (32.97±0.33)/(8.93±0.09), (37.34±0.29)/(9.75±0.08), respectively during the insertion of the suction tube from the suction port of the closed extension tube, and the tidal volume (in L) was decreased from0.361±0.005, 0.592±0.003, 0.631±0.001, 0.642±0.007, 0.633±0.007, 0.626±0.08 to 0.335±0.005, 0.588±0.008, 0.631±0.002, 0.638±0.004, 0.628±0.004, 0.618±0.005.The maximum pressure change of the ventilation circuit was (0.83±0.27) cm H2O and the maximum tidal volume change was (0.008±0.006)L. The changes of airway pressure and tidal volume were significantly lower than those of ventilation circuit connected with open extension tube under the same pressure parameters. Conclusion: The connection of closed extension tube in mechanical ventilation circuit can reduce the airway leakage during artificial airway clearance, which is worthy of clinical recommendation. 目的: 探讨密闭式延长管在人工气道吸痰作业期间预防气道漏气的作用。 方法: 采用人工模肺连接呼吸机进行机械通气,呼吸回路中分别连接开放式延长管或密闭式延长管,呼吸机参数设置为16/5、20/6、24/7、28/8、32/9、36/10 cmH2O(1 cmH2O=0.098 kPa)六组递增压力,观察吸痰作业期间气道压力和潮气量的变化。 结果: (1)呼吸机参数设置在16/5~36/10 cmH2O之间,连接开放式延长管机械通气回路的气道压力分别为(15.94±0.27)/(4.81±0.04)、(20.09±0.23)/(6.05±0.16)、(23.89±0.41)/(6.94±0.06)、(27.90±0.22)/(7.71±0.18)、(31.92±0.13)/(8.74±0.12)、(35.65±0.31)/(9.72±0.07)cmH2O;相同参数下连接密闭式延长管机械通气回路的气道压力分别为(16.36±0.06)/(4.85±0.04)、(20.54±0.26)/(6.44±0.12)、(24.36±0.24)/(7.01±0.33)、(28.69±0.25)/(8.07±0.08)、(32.97±0.33)/(8.93±0.09)、(37.34±0.29)/(9.75±0.08)cmH2O。连接开放式延长管机械通气回路的气道压力均低于连接密闭式延长管的(均P<0.05);随呼吸机设置压力的增加,两种延长管气道压力的差值逐渐增加,当呼吸机设置最高吸气压力为36 cmH2O时,差值达到1.69 cmH2O。(2)机械通气回路连接开放式延长管,吸痰作业拔插口开放时,通气回路的气道压力分别从(15.94±0.27)/(4.81±0.04)、(20.09±0.23)/(6.05±0.16)、(23.89±0.41)/(6.94±0.06)、(27.90±0.22)/(7.71±0.18)、(31.92±0.13)/(8.74±0.12)、(35.65±0.31)/(9.72±0.07)cmH2O下降至(13.42±0.4)/(3.15±0.14)、(16.81±0.6)/(4.30±0.14)、(20.22±0.5)/(5.48±0.45)、(23.73±1.4)/(6.25±0.22)、(24.78±0.7)/(7.13±0.21)、(20.83±0.4)/(6.61±0.19)cmH2O,(均P<0.05);潮气量分别从(0.328±0.004)、(0.580±0.012)、(0.621±0.003)、(0.626±0.003)、(0.615±0.003)、(0.603±0.002)L下降到(0.272±0.008)、(0.416±0.051)、(0.487±0.047)、(0.396±0.116)、(0.507±0.022)、(0.508±0.079)L(均P<0.05);随呼吸机设置的压力增加,通气回路的压力和潮气量下降幅度逐渐增加,当呼吸机设置参数为36/10 cmH2O时,吸气压力下降幅度高达(14.82±0.51)cmH2O;潮气量下降幅度最高时可达(0.164±0.021)L。(3)相同呼吸机参数下从处于开放状态的延长管插拔口插入吸痰管时,通气回路的气道压力分别上升为(15.70±0.23)/(4.80±0.33)、(19.01±0.81)/(5.71±0.34)、(22.27±0.62)/(6.85±0.44)、(25.35±2.09)/(7.94±0.16)、(28.38±0.46)/(8.96±0.23)、(33.34±0.71)/(9.71±0.25)cmH2O,潮气量分别为(0.340±0.016)、(0.563±0.020)、(0.571±0.030)、(0.556±0.026)、(0.514±0.021)、(0.512±0.031)L。气道压力和潮气量较插拔口开放状态时增加,但仍低于插拔口处于关闭状态的(均P<0.05),与插拔口处于关闭状态比较,压力下降幅度最大为(3.53±0.46)cmH2O,潮气量下降幅度最大为(0.101±0.011)L。(4)呼吸回路连接密闭式延长管,压力参数设置在16/5~36/10 cmH2O之间,从密闭式延长管的插拔口插入吸痰管期间,气道压力分别从(16.26±0.04)/(4.85±0.04)、(20.74±0.15)/(6.42±0.11)、(25.09±0.31)/(7.10±0.13)、(29.38±0.24)/(8.17±0.09)、(33.80±0.16)/(9.02±0.17)、(37.89±0.19)/(9.83±0.07)cmH2O下降到(16.36±0.06)/(4.85±0.04)、(20.54±0.26)/(6.44±0.12)、(24.36±0.24)/(7.01±0.33)、(28.69±0.25)/(8.07±0.08)、(32.97±0.33)/(8.93±0.09)、(37.34±0.29)/(9.75±0.08)cmH2O,潮气量分别从(0.361±0.005)、(0.592±0.003)、(0.631±0.001)、(0.642±0.007)、(0.633±0.007)、(0.626±0.08)L下降到(0.335±0.005)、(0.588±0.008)、(0.631±0.002)、(0.638±0.004)、(0.628±0.004)、(0.618±0.005)L。气道压力下降幅度最大为(0.83±0.27)cmH2O,潮气量下降幅度最大为(0.008±0.006)L,气道压力和潮气量的下降值均明显低于相同压力参数下通气回路连接开放式延长管的水平(均P<0.05)。 结论: 机械通气回路中连接密闭式延长管可降低人工气道吸痰作业期间气道漏气情况,值得临床推荐使用。.