电子口罩减少支气管镜操作中飞沫播散
本研究评估一次性电子口罩在支气管镜操作中减少气溶胶颗粒的保护能力,并前瞻性比较佩戴电子口罩与历史对照在呼吸不良事件及并发症上的差异。可视化与定量实验显示电子口罩显著减少气溶胶颗粒量(p=0.011),佩戴电子口罩未显著增加需>5 L/min氧气维持>90%血氧饱和度的发生率(p=0.959),但为EtCO2升高>45 mm Hg的显著因素(p=0.026),并发症无显著差异(5.8% vs. 2.5%,p=0.395)。
要点
- 电子口罩显著减少支气管镜操作中气溶胶颗粒量(p=0.011)。
- 佩戴电子口罩未显著增加需高流量氧维持血氧饱和度的发生率。
- 电子口罩是EtCO2升高>45 mm Hg的显著影响因素(p=0.026)。
核心结论
- 支气管镜操作中佩戴电子口罩显著减少气溶胶颗粒产生,且不增加并发症。
研究设计与样本
- 排除2-4期慢阻肺患者,进行气溶胶颗粒可视化与定量实验,并前瞻性评估佩戴电子口罩与历史对照的呼吸不良事件及并发症。
主要发现
- 可视化实验中无电子口罩时产生的气溶胶颗粒超过有电子口罩时的一万倍以上。
- 电子口罩组气溶胶颗粒体积较无电子口罩组显著降低(p=0.011)。
- 多因素logistic回归显示佩戴电子口罩对需>5 L/min氧气维持>90%血氧饱和度的发生率无显著影响(p=0.959)。
- 佩戴电子口罩是EtCO2升高>45 mm Hg的显著因素(p=0.026)。
- 电子口罩组与对照组并发症发生率无显著差异(5.8% vs. 2.5%,p=0.395)。
临床怎么用
- 支气管镜操作中可考虑为患者佩戴一次性电子口罩以减少气溶胶播散,但需关注EtCO2升高风险。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND AND OBJECTIVE: Bronchoscopy is an airborne particle-generating procedure. However, few methods for safe bronchoscopy have been developed. To reduce airborne particles during bronchoscopy, we created an 'e-mask', which is a simple, disposable mask for patients. Our objective was to evaluate the e-mask's protective ability against airborne particles and to assess respiratory adverse events and complications. METHODS: Patients with stage 2-4 chronic obstructive pulmonary disease were excluded. We performed visualization and quantifying experiments on airborne particles with and without the e-mask. We prospectively evaluated whether wearing the e-mask during bronchoscopy was associated with the incidence of patients requiring >5 L/min oxygen to maintain >90% oxygen saturation, and patients with >45 mm Hg end-tidal carbon dioxide (EtCO2 ) elevation, in addition to complications, compared to historical controls. RESULTS: In the visualization experiment, more than ten thousand times of airborne particles were generated without the e-mask than with the e-mask. The volume of airborne particles was significantly reduced with the e-mask, compared to that without the e-mask (p = 0.011). Multivariate logistic regression analysis revealed that wearing the e-mask had no significant effect on the incidence of patients requiring >5 L/min oxygen to maintain >90% oxygen saturation, (p = 0.959); however, wearing the e-mask was a significant factor in >45 mm Hg EtCO2 elevation (p = 0.026). No significant differences in complications were observed between the e-mask and control groups (5.8% vs. 2.5%, p = 0.395). CONCLUSION: Wearing the e-mask during bronchoscopy significantly reduced the generation of airborne particles during bronchoscopy without increasing complications.
引用
- Ito T, Okachi S, Sato K, et al. Prevention of droplet dispersal with 'e-mask': A new daily use endoscopic mask during bronchoscopy. Respirology. 2022. PMID: 35781913. doi:10.1111/resp.14321
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。