产程中胎儿头皮血采样乳酸与pH比较的临床单中心研究:迈向临床实施
该前瞻性观察性单中心研究纳入263例产程中胎儿头皮血样本,比较StatStrip® Lactate 2测乳酸与iSTAT®测pH的相关性并推导乳酸截断值;乳酸与pH呈中度负相关(r=-0.53),乳酸≥5.6 mmol/L对pH<7.20的灵敏度78.1%、特异度79.2%。
要点
- 胎儿头皮血乳酸与pH呈中度负相关(r=-0.53)。
- 乳酸识别胎儿头皮血pH<7.20的ROC曲线AUC为0.88(0.83–0.93)。
- 乳酸≥5.6 mmol/L对pH<7.20的灵敏度78.1%、特异度79.2%。
核心结论
- 研究证实StatStrip® Lactate 2可作为iSTAT®测pH的合适替代分析仪用于胎儿头皮血乳酸测定。
- 尽管乳酸与pH仅中度相关,可提出乳酸≥5.6 mmol/L用于临床决策。
研究设计与样本
- 观察性前瞻性单中心研究,8个月内纳入263例胎儿头皮血样本。
- 采集胎儿头皮血后先以iSTAT®测pH,再以StatStrip® Lactate 2测乳酸,临床决策仅基于pH结果。
主要发现
- 胎儿头皮血乳酸与pH呈中度负向线性相关(r=-0.53)。
- 乳酸识别胎儿头皮血pH<7.20的ROC曲线AUC为0.88(0.83–0.93),提示诊断性能良好。
- 乳酸≥5.6 mmol/L对胎儿头皮血pH<7.20的灵敏度为78.1%(60.0–90.7%),特异度为79.2%(73.4–84.3%),为灵敏度与特异度的最佳平衡。
关键数值与推荐
- 乳酸≥5.6 mmol/L对胎儿头皮血pH<7.20的灵敏度78.1%(60.0–90.7%)、特异度79.2%(73.4–84.3%)。
- 乳酸识别胎儿头皮血pH<7.20的ROC曲线AUC为0.88(0.83–0.93)。
临床怎么用
- 作者认为乳酸≥5.6 mmol/L可被提出用于临床决策。
- 作者指出该知识增强了其妇科医生对这一相对新方法的信心,并为在胎儿头皮血中成功且普遍实施乳酸测定、最终无需测pH铺平道路。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
Lactate and pH in fetal blood samples are moderately correlated (r = −0.53). For the StatStrip® Lactate 2, we propose a lactate cutoff value of ≥5.6 mmol/L based on the most optimal balance between sensitivity (78.1% [60.0–90.7%]) and specificity (79.2% [73.4–84.3%]). Abstract Introduction Since 60 years, pH measurement in fetal scalp blood samples (FBS) is used to diagnose fetal hypoxia during labor, although the last three decades lactate measurement in FBS has been increasingly used as a safe alternative with a lower chance of failure. However, studies demonstrating the direct relation between lactate and pH in FBS with the currently used measurement systems are lacking. Therefore, clinical acceptance and implementation is impeded. This study investigated the relation between lactate and pH in FBS and derived an appropriate lactate cutoff value. Material and Methods This observational prospective single-center study included 263 FBS in the period of 8 months. After collection of FBS, pH was assessed first (iSTAT®), after which lactate was assessed (StatStrip® Lactate 2). Clinical decisions were based on pH results only. Our primary objective was to assess the correlation between lactate and pH in FBS and to establish a lactate cutoff value for ruling out a low Apgar score at 5 minutes and neonatal acidosis, with corresponding sensitivity and specificity as secondary outcomes. Diagnostic performance of lactate was assessed by using the area under the curve (AUC) of the receiver operation characteristic curve (ROC curve), from which cutoff values were derived. Results In FBS, lactate and pH showed a moderate inverse linear correlation (r = −0.53). The AUC of the ROC curve was 0.88 (0.83–0.93), indicating a good diagnostic performance of lactate to identify an abnormal pH (pH <7.20 in FBS). A cutoff value for lactate of ≥5.6 mmol/L seemed to have the most optimal balance between sensitivity (78.1% [60.0–90.7%]) and specificity (79.2% [73.4–84.3%]) for a pH <7.20 in FBS. Conclusions This study confirmed that the StatStrip® Lactate 2 is a suitable alternative analyzer for measurement of lactate compared with measurement of pH with the iSTAT in FBS. Despite only a moderate correlation between lactate and pH, a lactate cutoff value of ≥5.6 mmol/L could be proposed for clinical decision-making. This knowledge increased confidence of our gynecologists in this relatively new method and paved the way for successful and universal implementation of lactate in FBS, eventually without measuring pH.
引用
- Acta Obstetricia et Gynecologica Scandinavica:Toward implementation of lactate compared with pH in fetal scalp blood sampling during labor: A clinical comparative single‐center study
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。