新型邮寄精液分析系统的开发验证及即时与延迟检测的相关性
本研究开发并验证了一种新型邮寄精液分析系统,对104份正常精液样本进行1小时检测后,在52小时内重复检测4次。结果显示精子浓度保持稳定,活力每小时下降0.39%,正常形态每小时下降0.1%,1小时与延迟检测的活力(相关系数0.87)和形态(相关系数0.82)高度相关。
要点
- 邮寄精液分析系统在52小时内精子浓度稳定,活力与形态呈线性下降。
- 1小时与延迟检测的活力相关系数0.87,形态相关系数0.82。
- 该系统可简化精液获取流程,降低男性生育评估的早期障碍。
核心结论
- 新型邮寄精液分析系统在1小时与延迟检测间显示强相关性,可用于临床生育评估。
研究设计与样本
- 前瞻性队列研究,纳入正常精液男性的射精样本104份,验证阶段进行1小时精液分析。
主要发现
- 在长达52小时的观察和每份精液后续4次检测中,精子浓度保持稳定。
- 精子活力每小时下降0.39%,正常形态每小时下降0.1%。
- 1小时检测与计算值的活力相关系数为0.87,形态相关系数为0.82。
关键数值与推荐
- 世界卫生组织精液分析参数用于评估。
- 该系统获Clinical Laboratory Improvement Amendments批准。
临床怎么用
- 该检测可用于临床实践中的生育评估,提高精液分析获取的便捷性、舒适性和效率。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: To develop and validate a novel, mail-in semen analysis (SA) system. DESIGN: Prospective cohort. SETTING: Not applicable. PATIENT(S): Ejaculates from normospermic men. INTERVENTION(S): One-hour SA, then repeat SAs (on same ejaculate) over 52 hours using a novel technique for maintaining sperm viability. MAIN OUTCOME MEASURE(S): World Health Organization SA parameters. RESULT(S): One-hour SA on 104 ejaculates in the validation phase of the study demonstrated normal semen parameters. With up to 52 hours of observation and four subsequent SA measurements/ejaculate, concentration remained stable, motility decreased by 0.39%/h, and normal morphology decreased by 0.1%/h. Measured 1-hour and calculated motility (correlation coefficients 0.87) and morphology (correlation coefficients 0.82) strongly were correlated. CONCLUSION: This novel, mail-in, Clinical Laboratory Improvement Amendments-approved SA testing system demonstrates a strong degree of correlation between 1-hour and delayed SA testing. Given the linear motility and morphology decrease and stability of sperm concentration, this test may be used in clinical practice to evaluate semen quality for fertility evaluations. Furthermore, this approach significantly improves the ease, comfort, and efficiency of obtaining a SA, likely breaking down early barriers to accessing successfully a male fertility evaluation.
引用
- Samplaski MK, Falk O, Honig S, et al. Development and validation of a novel mail-in semen analysis system and the correlation between one hour and delayed semen analysis testing. Fertil Steril. 2021. PMID: 33423785. doi:10.1016/j.fertnstert.2020.10.047
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。