文献速递母胎医学

智能气管封堵装置在膈疝胎羊模型中的安全性与有效性

Safety and efficacy of smart tracheal occlusion device in diaphragmatic hernia lamb model.
PubMed 文献 · Ultrasound Obstet Gynecol · 2021-01-01

在胎羊膈疝模型中,于孕97天用Smart-TO球囊行胎儿镜下气管封堵,孕116天经3.0T磁共振使球囊无创放气。与未封堵的膈疝胎羊相比,封堵组出生时肺体重比显著更高,但肺泡增大且气道复杂度降低;气管组织学改变轻微但较未封堵组更明显。

要点

  • Smart-TO球囊可经磁共振无创放气并自行排出气道。
  • 封堵后肺体重比显著升高,但肺泡增大且终末细支气管密度降低。
  • 气管组织学改变轻微,但封堵组较未封堵组更明显。

核心结论

  • 在膈疝胎羊中,使用Smart-TO球囊行气管封堵有效且安全,封堵可无创逆转,放气后的完整球囊可自行从胎儿上气道排出。

研究设计与样本

  • 在孕70天(足月145天)创建胎羊膈疝模型,分为继续妊娠至足月的DH组和孕97天用Smart-TO球囊行胎儿镜下气管封堵的DH+TO组。
  • 孕116天经超声确认球囊存在后,母羊绕3.0T磁共振扫描仪行走使球囊放气,随后立即超声确认。
  • 足月处死取胎,以肺体重比和肺形态计量评估有效性,以气管形态计量和放气后球囊位置评估安全性。
  • DH组6只胎羊,DH+TO组7只胎羊。

主要发现

  • DH+TO组出生时肺体重比显著高于对照组(1.90(IQR 1.43-2.55)vs 1.07(IQR 0.93-1.46);P=0.005)。
  • DH+TO组平均线性截距显著增大(47.5(IQR 45.6-48.1)vs 41.9(IQR 38.8-46.1)μm;P=0.03)。
  • DH+TO组平均终末细支气管密度显著降低(1.56(IQR 1.0-1.81)vs 2.23(IQR 2.14-2.40)br/mm²;P=0.005)。
  • 两组气管组织学改变均轻微,但DH+TO组气管评分高于未封堵组(2(IQR 1-3)vs 0(0-1);P=0.03)。
  • 所有球囊均成功放气并自行从气道排出。

临床怎么用

  • 该动物模型结果支持Smart-TO球囊作为可无创逆转的胎儿气管封堵方案进一步研究。
  • 临床转化前需注意封堵后肺泡增大和气道复杂度降低的形态学改变。

局限性

  • 样本量小(DH组6只、DH+TO组7只胎羊)。
  • 为胎羊动物模型,结果不能直接外推至人类胎儿。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: To investigate the efficacy and safety of the 'smart' tracheal occlusion (Smart-TO) device in fetal lambs with diaphragmatic hernia (DH). METHODS: DH was created in fetal lambs on gestational day 70 (term, 145 days). Fetuses were allocated to either pregnancy continuation until term (DH group) or fetoscopic endoluminal tracheal occlusion (TO), performed using the Smart-TO balloon on gestational day 97 (DH + TO group). On gestational day 116, the presence of the balloon was confirmed on ultrasound, then the ewe was walked around a 3.0-Tesla magnetic resonance scanner for balloon deflation, which was confirmed by ultrasound immediately afterwards. At term, euthanasia was performed and the fetus retrieved. Efficacy of occlusion was assessed by the lung-to-body-weight ratio (LBWR) and lung morphometry. Safety parameters included tracheal side effects assessed by morphometry and balloon location after deflation. The unoccluded DH lambs served as a comparator. RESULTS: Six fetuses were included in the DH group and seven in the DH + TO group. All balloons deflated successfully and were expelled spontaneously from the airways. In the DH + TO group, in comparison to controls, the LBWR at birth was significantly higher (1.90 (interquartile range (IQR), 1.43-2.55) vs 1.07 (IQR, 0.93-1.46); P = 0.005), while on lung morphometry, the alveolar size was significantly increased (mean linear intercept, 47.5 (IQR, 45.6-48.1) vs 41.9 (IQR, 38.8-46.1) μm; P = 0.03); whereas airway complexity was lower (mean terminal bronchiolar density, 1.56 (IQR, 1.0-1.81) vs 2.23 (IQR, 2.14-2.40) br/mm2 ; P = 0.005). Tracheal changes on histology were minimal in both groups, but more noticeable in fetal lambs that underwent TO than in unoccluded lambs (tracheal score, 2 (IQR, 1-3) vs 0 (0-1); P = 0.03). CONCLUSIONS: In fetal lambs with DH, TO using the Smart-TO balloon is effective and safe. Occlusion can be reversed non-invasively and the deflated intact balloon expelled spontaneously from the fetal upper airways. © 2020 International Society of Ultrasound in Obstetrics and Gynecology.

来源:Ultrasound Obstet Gynecol,查看原文。版权归原作者所有。

引用

  1. Basurto D, Sananès N, Bleeser T, et al. Safety and efficacy of smart tracheal occlusion device in diaphragmatic hernia lamb model. Ultrasound Obstet Gynecol. 2021. PMID: 33012007. doi:10.1002/uog.23135

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。