自动三维超声成像技术用于胎儿头部生物测量的准确性

Accuracy of automated three-dimensional ultrasound imaging technique for fetal head biometry.
PubMed 文献 · Ultrasound Obstet Gynecol · 2021-01-01

本研究为前瞻性观察性研究,纳入143例孕18+0至22+6周接受常规解剖扫描的孕妇,比较人工智能工具SonoCNS® Fetal Brain自动与人工测量胎儿双顶径、头围、小脑横径、小脑延髓池和侧脑室后角的准确性。自动测量双顶径和头围信度良好(ICC 0.80-0.88),小脑横径、小脑延髓池和侧脑室后角信度差至中等(ICC 0.23-0.50)。作者认为该技术有前景,但需进一步优化后方可纳入常规超声方案。

要点

  • 自动测量双顶径和头围与人工测量信度良好,ICC为0.80-0.88。
  • 自动测量小脑横径、小脑延髓池和侧脑室后角信度差至中等,ICC为0.23-0.50。
  • 作者认为需进一步优化该技术后方可纳入常规超声方案。

核心结论

  • SonoCNS自动三维超声可可靠识别和测量胎儿双顶径和头围,但识别和测量小脑横径、小脑延髓池和侧脑室后角的一致性较差。
  • 作者认为结果令人鼓舞,但在将该技术纳入常规超声方案前需进一步优化。

研究设计与样本

  • 前瞻性观察性研究,纳入143例孕18+0至22+6周接受常规解剖扫描的孕妇。
  • 每例患者由一名超声技师和一名医师分别进行连续两次胎儿头部超声检查,各操作者先手工测量双顶径、头围、小脑横径、小脑延髓池和侧脑室后角,再使用基于人工智能的工具SonoCNS® Fetal Brain进行自动测量。
  • 两名操作者重复自动测量方法直至单次扫描获取全部五项测量,最多尝试三次。
  • 采用组内相关系数(ICC)按操作者类型比较自动与手工测量的准确性,并考虑患者和超声特征。

主要发现

  • 孕妇中位体重指数为24.0 kg/m2(四分位距22.5-26.8 kg/m2),中位皮下厚度为1.6 cm(四分位距1.3-2.0 cm)。
  • 15例(10%)患者至少有一次剖宫产史,17例(12%)有其他腹部手术史,78例(55%)为前壁胎盘。
  • 超声技师首次、第二次和第三次尝试成功获取自动测量的比例分别为70%、22%和3%;医师分别为76%、16%和3%。
  • 自动算法在超声技师和医师扫描中分别有6例(4%)和7例(5%)未能正确识别和测量全部五项结构。
  • 自动测量与手工测量相比,双顶径和头围的ICC为0.80-0.88,信度良好;小脑横径、小脑延髓池和侧脑室后角的ICC为0.23-0.50,信度差至中等。

局限性

  • 原文结论指出,在将该技术纳入常规超声方案前需进一步优化自动技术。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: To evaluate the accuracy of an automated three-dimensional (3D) ultrasound technique for fetal intracranial measurements compared with manual acquisition. METHODS: This was a prospective observational study of patients presenting for routine anatomical survey between 18 + 0 and 22 + 6 weeks' gestation. After providing informed consent, each patient underwent two consecutive ultrasound examinations of the fetal head, one by a sonographer and one by a physician. Each operator obtained manual measurements of the biparietal diameter (BPD), head circumference (HC), transcerebellar diameter (TCD), cisterna magna (CM) and posterior horn of the lateral ventricle (Vp), followed by automated measurements of these structures using an artificial intelligence-based tool, SonoCNS® Fetal Brain. Both operators repeated the automated approach until all five measurements were obtained in a single sweep, up to a maximum of three attempts. The accuracy of automated measurements was compared with that of manual measurements using intraclass correlation coefficients (ICC) by operator type, accounting for patient and ultrasound characteristics. RESULTS: One hundred and forty-three women were enrolled in the study. Median body mass index was 24.0 kg/m2 (interquartile range (IQR), 22.5-26.8 kg/m2 ) and median subcutaneous thickness was 1.6 cm (IQR, 1.3-2.0 cm). Fifteen (10%) patients had at least one prior Cesarean delivery, 17 (12%) had other abdominal surgery and 78 (55%) had an anterior placenta. Successful acquisition of the automated measurements was achieved on the first, second and third attempts in 70%, 22% and 3% of patients, respectively, by sonographers and in 76%, 16% and 3% of cases, respectively, by physicians. The automated algorithm was not able to identify and measure all five structures correctly in six (4%) and seven (5%) patients scanned by the sonographers and physicians, respectively. The ICCs reflected good reliability (0.80-0.88) of the automated compared with the manual approach for BPD and HC and poor to moderate reliability (0.23-0.50) for TCD, CM and Vp. Fetal lie, head position, placental location, maternal subcutaneous thickness and prior Cesarean section were not associated with the success or accuracy of the automated technique. CONCLUSIONS: Automated 3D ultrasound imaging of the fetal head using SonoCNS reliably identified and measured BPD and HC but was less consistent in accurately identifying and measuring TCD, CM and Vp. While these results are encouraging, further optimization of the automated technology is necessary prior to incorporation of the technique into routine sonographic protocols. © 2020 International Society of Ultrasound in Obstetrics and Gynecology.