新型装置实现胎儿内镜下气管封堵的无创逆转:离体与在体研究
本研究在模拟宫内环境及胎羊模型中评估Smart-TO球囊的可逆性、局部副作用与封堵效果。离体模拟中,1.5-T磁共振暴露使多数球囊放气,仅坐轮椅体位有1例三次暴露后仍未放气;在体胎羊中Smart-TO球囊均成功放气,肺体重比显著高于未封堵对照且与Goldbal2相当,气管组织学改变轻微。
要点
- Smart-TO球囊在1.5-T磁共振暴露下可放气,离体模拟中坐轮椅体位有1例三次暴露后仍未放气。
- 胎羊在体实验中Smart-TO球囊均成功放气,肺体重比显著高于未封堵对照。
- Smart-TO与Goldbal2在肺形态计量、气管周径及气管组织学改变上无差异。
核心结论
- 在模拟宫内环境中,胎儿处于不同体位时暴露于1.5-T磁共振系统磁场可有效使Smart-TO球囊放气,仅母亲坐轮椅体位时出现1例失败。
- 在健康胎羊中,Smart-TO球囊的封堵效果与临床标准Goldbal2系统相当,且局部副作用有限。
研究设计与样本
- 离体研究:在高保真模拟器中经视频镜插入气管球囊,将胎儿模型置于1-L充水球囊内模拟羊膜腔,由操作者持于腹部前方,模拟不同胎儿与母亲体位。
- 在体研究:胎羊于妊娠第95天(足月145天)行气管封堵,Goldbal2组n=5、Smart-TO组n=5,另6只未封堵胎羊为对照。
- 妊娠第116天经气管镜评估球囊,Goldbal2经穿刺放气、Smart-TO经磁共振暴露放气,安乐死后评估肺体重比、肺形态计量、气管周径及气管组织学评分。
主要发现
- 离体模拟中,母亲站立位且模型高度95 cm(n=32)、55 cm(n=8)或125 cm(n=8)时,以及母亲「躺在担架上」(n=8)时,单次磁共振暴露后100%球囊放气。
- 离体模拟中,母亲「坐在轮椅上」时8个球囊中3个(37.5%)首次暴露未放气,其中2个第二次暴露后放气,1个第三次暴露后仍充气。
- 在体实验中所有Smart-TO球囊均成功放气;Smart-TO组胎羊肺体重比显著高于未封堵对照组,与Goldbal2组相当。
- 两种球囊在肺形态计量和气管周径上无差异,气管组织学均显示轻微改变。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: One of the drawbacks of fetal endoscopic tracheal occlusion (FETO) for congenital diaphragmatic hernia is the need for a second invasive intervention to re-establish airway patency. The 'Smart-TO' device is a new balloon for FETO that deflates spontaneously when placed in a strong magnetic field, therefore overcoming the need for a second procedure. The safety and efficacy of this device have not yet been demonstrated. The aim of this study was to investigate the reversibility, local side effects and occlusiveness of the Smart-TO balloon, both in a simulated in-utero environment and in the fetal lamb model. METHODS: First, the reversibility of tracheal occlusion by the Smart-TO balloon was tested in a high-fidelity simulator. Following videoscopic tracheoscopic balloon insertion, the fetal mannequin was placed within a 1-L water-filled balloon to mimic the amniotic cavity. This was held by an operator in front of their abdomen, and different fetal and maternal positions were simulated to mimic the most common clinical scenarios. Following exposure to the magnetic field generated by a 1.5-T magnetic resonance (MR) machine, deflation of the Smart-TO balloon was assessed by tracheoscopy. In cases of failed deflation, the mannequin was reinserted into a water-filled balloon for additional MR exposure, up to a maximum of three times. Secondly, reversibility, occlusiveness and local effects of the Smart-TO balloon were tested in vivo in fetal lambs. Tracheal occlusion was performed in fetal lambs on gestational day 95 (term, 145 days), either using the balloon currently used in clinical practice (Goldbal2) (n = 5) or the Smart-TO balloon (n = 5). On gestational day 116, the presence of the balloon was assessed by tracheoscopy. Deflation was performed by puncture (Goldbal2) or MR exposure (Smart-TO). Six unoccluded fetal lambs served as controls. Following euthanasia, the lung-to-body-weight ratio (LBWR), lung morphometry and tracheal circumference were assessed. Local tracheal changes were measured using a hierarchical histologic scoring system. RESULTS: Ex vivo, Smart-TO balloon deflation occurred after a single MR exposure in 100% of cases in a maternal standing position with the mannequin at a height of 95 cm (n = 32), 55 cm (n = 8) or 125 cm (n = 8), as well as when the maternal position was 'lying on a stretcher' (n = 8). Three out of eight (37.5%) balloons failed to deflate at first exposure when the maternal position was 'sitting in a wheelchair'. Of these, two balloons deflated after a second MR exposure, but one balloon remained inflated after a third exposure. In vivo, all Smart-TO balloons deflated successfully. The LBWR in fetal lambs with tracheal occlusion by a Smart-TO balloon was significantly higher than that in unoccluded controls, and was comparable with that in the Goldbal2 group. There were no differences in lung morphometry and tracheal circumference between the two balloon types. Tracheal histology showed minimal changes for both balloons. CONCLUSIONS: In a simulated in-utero environment, the Smart-TO balloon was effectively deflated by exposure of the fetus in different positions to the magnetic field of a 1.5-T MR system. There was only one failure, which occurred when the mother was sitting in a wheelchair. In healthy fetal lambs, the Smart-TO balloon is as occlusive as the clinical standard Goldbal2 system and has only limited local side effects. © 2020 International Society of Ultrasound in Obstetrics and Gynecology.