文献速递产科母胎医学

高危剖宫产术中采用电生物电抗或动脉波形分析测量心输出量:一致性评估

Cardiac output measurements during high-risk Cesarean section using electrical bioreactance or arterial waveform analysis: assessment of agreement.
PubMed 文献 · Ultrasound Obstet Gynecol · 2019-07-11

本研究在高危血流动力学不稳定的剖宫产产妇中,同时使用有创LiDCOrapid™和无创NICOM®设备测量心输出量(CO),分析10例产妇的307对测量数据。结果显示两种设备CO测量的平均偏差为3.05 L/min(95%CI 1.89至4.21),一致性界限为-1.58至7.68 L/min,一致性间隔9.26 L/min,一致性:耐受性指数(ATI)为2.3。结论认为两设备测量差异大,临床不可互换,一致性低(ATI>2),在大出血时存在高临床误分类风险。

要点

  • 高危剖宫产术中LiDCOrapid与NICOM测CO平均偏差3.05 L/min(95%CI 1.89至4.21)。
  • 两设备一致性界限-1.58至7.68 L/min,一致性间隔9.26 L/min,ATI为2.3。
  • ATI>2提示一致性不可接受,两设备临床不可互换,大出血时误分类风险高。

核心结论

  • 高危剖宫产术中LiDCOrapid与NICOM测量CO的平均差异大,临床不应视为可互换。
  • 两设备CO测量一致性低(ATI>2),在大出血时存在高临床误分类风险。

研究设计与样本

  • 研究在高危血流动力学不稳定的剖宫产产妇中,按标准化技术同时使用有创LiDCOrapid™和无创NICOM®设备进行术中血流动力学测量。
  • 共纳入10例产妇,分析307对同时测量的CO数据。
  • 采用Bland-Altman图和一致性:耐受性指数(ATI)评估两设备CO测量的一致性,并对重复测量使用百分位数进行一致性分析。

主要发现

  • LiDCOrapid与NICOM设备CO测量的平均偏差为3.05 L/min(95%CI 1.89至4.21)。
  • 一致性界限范围为-1.58 L/min(95%CI -4.47至-0.14)至7.68 L/min(95%CI 6.24至10.56)。
  • 一致性间隔为9.26 L/min,对应ATI为2.3。

关键数值与推荐

  • ATI>2提示两设备CO测量一致性不可接受。
  • 平均偏差3.05 L/min(95%CI 1.89至4.21),一致性间隔9.26 L/min。

临床怎么用

  • 作者认为LiDCOrapid与NICOM设备在高危剖宫产术中测量的CO不可临床互换。
  • 作者指出两设备一致性低(ATI>2),在大出血期间存在高临床误分类风险。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: Maternal hemodynamics change significantly during Cesarean section complicated by massive hemorrhage or severe hypertensive disease. Cardiac output (CO) monitoring aids early, goal-directed hemodynamic therapy. The aim of this study was to record hemodynamic changes observed during Cesarean section in pregnancies at high risk of hemodynamic instability, using invasive (LiDCOrapid™) and non-invasive (NICOM®) devices, and to assess agreement between the two devices in measuring CO. METHODS: Simultaneous intraoperative hemodynamic measurements were taken using the LiDCOrapid and NICOM devices, following standardized techniques, in women at high risk of hemodynamic instability undergoing Cesarean section. Agreement in CO measurements between the two devices was assessed using Bland-Altman plots and the agreement:tolerability index (ATI). Agreement analyses were performed for repeated measures in subjects, using centiles. RESULTS: From 10 women, 307 paired measurements were analyzed. Mean bias (defined as the mean difference in CO measurements between the LiDCOrapid and NICOM devices) was 3.05 (95% CI, 1.89 to 4.21) L/min. Limits of agreement ranged from -1.58 (95% CI, -4.47 to -0.14) to 7.68 (95% CI, 6.24 to 10.56) L/min. The resulting agreement interval was 9.26 L/min which returned an ATI of 2.3. CONCLUSIONS: There are large mean differences between CO measurements obtained during Cesarean section using the LiDCOrapid and NICOM hemodynamic monitors in pregnant women at high risk of hemodynamic instability, indicating that they should not be considered interchangeable clinically. There is an unacceptably low level of agreement (ATI > 2) in CO measurements between the devices, conferring a high risk of clinical misclassification during massive hemorrhage. Copyright © 2018 ISUOG. Published by John Wiley & Sons Ltd.

来源:Ultrasound Obstet Gynecol,查看原文。版权归原作者所有。

引用

  1. Gutierrez J, Perry H, Columb M, et al. Cardiac output measurements during high-risk Cesarean section using electrical bioreactance or arterial waveform analysis: assessment of agreement. Ultrasound Obstet Gynecol. 2019. PMID: 30302868. doi:10.1002/uog.20142

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。