Aveta宫腔组织去除系统在标准化肌瘤模型中的切除效率比较
这项前瞻性多中心台式研究使用模拟致密肌瘤机械阻力的合成模型,比较Aveta Flex与三种成熟宫腔旋切设备的切除效率。91名妇产科医生参与,共220次设备观察,Aveta Flex的最小二乘均值切除率为0.506 g/min,显著高于MyoSure Reach、MyoSure XL和TruClear DTSP。
要点
- 研究采用合成肌瘤模型,模拟高度纤维化或部分钙化肌瘤的机械阻力。
- 切除效率差异可能影响手术时间、完全切除概率及宫腔镜肌瘤切除术流程。
核心结论
- 在标准化合成肌瘤模型中,Aveta Flex宫腔组织去除系统的切除效率显著高于现有宫腔旋切设备。
研究设计与样本
- 前瞻性、多中心、重复测量、不完全区组设计的台式比较研究,91名常规行宫腔镜肌瘤切除术的妇产科执业医生参与,共220次设备观察。
主要发现
- Aveta Flex(n=91)最小二乘均值切除率为0.506 g/min,MyoSure Reach(n=56)为0.184 g/min,MyoSure XL(n=37)为0.274 g/min,TruClear DTSP(n=36)为0.389 g/min。
临床怎么用
- 在宫腔镜肌瘤切除术中,选择切除效率更高的设备可能有助于缩短手术时间、提高完全切除概率并优化流程,尤其需考虑液体亏欠阈值限制。
局限性
- 研究使用合成肌瘤类似物模型而非真实人体组织,结果外推至临床需谨慎。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
STUDY OBJECTIVE: To evaluate the resection efficiency of the Aveta hysteroscopic tissue removal system and compare its performance with three established hysteroscopic morcellation devices using a standardized synthetic fibroid model designed to simulate the mechanical resistance of dense leiomyoma tissue. STUDY DESIGN: Prospective, multicenter comparative bench study with repeated measures in an incomplete blocks design. SETTING: Simulated hysteroscopic fibroid resections under standardized conditions. PARTICIPANTS: Ninety-one Ob/Gyn practicing physicians routinely performing hysteroscopic myomectomy procedures participated in the study. INTERVENTIONS: Simulated hysteroscopic resections using 4 commercially available hysteroscopic tissue removal systems (Aveta Flex, MyoSure Reach, MyoSure XL, and TruClear Dense Tissue Shaver Plus (DTSP)) were performed in a synthetic fibroid analog tissue model with rheologic properties calibrated to simulate the mechanical resistance of highly fibrotic or partially calcified leiomyomas. Resection efficiency was quantified by measuring the mass of fibroid analog removed in g/min. MEASUREMENTS AND RESULTS: Ninety-one physicians participated and a total of 220 device observations were included: Aveta Flex (n = 91), MyoSure Reach (n = 56), MyoSure XL (n = 37), and TruClear DTSP (n = 36). Least squares mean resection rates were 0.506 g/min for Aveta Flex, 0.184 g/min for MyoSure Reach, 0.274 g/min for MyoSure XL, and 0.389 g/min for TruClear DTSP. Parametric and nonparametric analyses demonstrated significantly higher resection rates for Aveta Flex compared with MyoSure Reach, MyoSure XL, and TruClear DTSP (mixed-effects model p < .005; Skillings-Mack p < .0001). CONCLUSION: In this prospective, multicenter study using a standardized synthetic fibroid analog model, the Aveta Flex hysteroscopic tissue removal system demonstrated significantly higher resection efficiency compared with established hysteroscopic morcellation devices. Given the importance of procedural efficiency in operative hysteroscopy, particularly within the limitations imposed by fluid deficit thresholds, these findings may have implications for operative duration, likelihood of complete resection, and overall procedural workflow in hysteroscopic myomectomy.
引用
- Berman JM, Haimovich S, Levine DJ, et al. Comparative Resection Efficiency of a Hysteroscopic Tissue Removal System: A Prospective Multicenter Bench Study Using a Standardized Fibroid Model. J Minim Invasive Gynecol. 2026. PMID: 42468878. doi:10.1016/j.jmig.2026.07.012
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。