骶棘韧带固定术中锚钉脱位导致的术者与患者损伤:一项回顾性研究
该回顾性研究纳入2018年4月至2023年11月期间676例接受锚钉装置行骶棘韧带固定术的Ⅲ期及以上盆腔器官脱垂患者,其中657例使用经阴道网片、19例单独使用锚钉。锚钉脱位发生率为2.2%,脱位组体重指数和高血压患病率显著更高,并出现2例膀胱损伤和4例术者损伤。
要点
- 锚钉脱位发生率为2.2%,脱位组体重指数显著更高(p < 0.001)。
- 脱位组高血压患病率显著更高(p = 0.004),并发生2例膀胱损伤和4例术者损伤(p < 0.001)。
- 无脱位组失血量显著少于脱位组(p < 0.001),两组主客观治愈率均较高且相近。
核心结论
- 使用锚钉装置的骶棘韧带固定术安全有效,识别锚钉脱位相关风险可提高患者和术者的防范准备并减少并发症。
研究设计与样本
- 回顾性研究,在三级医疗中心纳入676例Ⅲ期及以上盆腔器官脱垂患者,均于2018年4月至2023年11月期间接受锚钉装置行骶棘韧带固定术。
- 其中657例使用经阴道网片(Surelift 463例,Calistar-S 194例),19例单独使用Anchorsure。
- 患者分为无脱位组(n = 649)和锚钉脱位组(n = 27),采用标准化机构方案,包括术前评估及基线、术后6个月和12个月经验证的中文问卷随访。
主要发现
- 锚钉脱位发生率为2.2%。
- 脱位组体重指数显著更高(p < 0.001),高血压患病率显著更高(p = 0.004)。
- 脱位组并发症包括2例膀胱损伤和4例术者损伤(包括手套撕裂和手指割伤)(p < 0.001)。
- 无脱位组失血量显著少于脱位组(p < 0.001)。
- 主客观治愈率两组均较高且相近:无脱位组分别为95.5%和94.4%,脱位组为90%;生活质量和性功能结局同样良好。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: To identify the factors contributing to tacker dislodgement and associated complications during sacrospinous ligament fixation (SSF) procedures. DESIGN: Retrospective study. SETTING: Tertiary medical center. POPULATION: Six hundred and seventy-six patients with stage ≥ 3 pelvic organ prolapse (POP) who underwent SSF with anchor-based devices (tackers) between April 2018 and November 2023. METHODS: Of these, 657 underwent SSF with transvaginal mesh (Surelift: 463; Calistar-S: 194), and 19 underwent SSF with Anchorsure alone. Patients were grouped into secure (n = 649) and dislodged tacker (n = 27) groups. Standardised institutional protocols, included pre-operative assessments, validated Chinese questionnaires at baseline and follow-up at 6 and 12 months postoperatively. MAIN OUTCOME MEASURES: Determine the occurrence of a tacker dislodge during application and the injury caused to the patient and the operator. RESULTS: Tacker dislodgement occurred in 2.2% of the cases. The dislodged group had a significantly higher BMI (p < 0.001) and prevalence of hypertension (p = 0.004). Complications in this group included two bladder injuries and four operator injuries, including glove tears and finger cuts (p < 0.001). The secure group demonstrated significantly less blood loss than the non-secure group (p < 0.001). Objective and subjective cure rates were high and comparable between the groups: 95.5% and 94.4% for the secure group and 90% for the dislodged group. The quality of life and sexual function outcomes were similarly favourable. CONCLUSION: SSF using a tacker-based device was safe and effective. Recognising the risks associated with tacker dislodgement can enhance preparedness and minimise complications for both patients and operators.