使用人工智能软件辅助识别与胎儿先天性心脏缺陷相关的超声表现
本研究回顾性构建了来自11个中心的200例胎儿超声检查数据集,其中100例至少含一项可疑表现。7名妇产科医生和7名母胎医学专家分别在有无AI辅助下阅片,评估识别可疑先天性心脏缺陷表现的能力。AI辅助显著提高了识别任何可疑表现的AUROC、敏感度和特异度,并缩短了解读时间、提高了信心评分。
要点
- AI辅助使识别任何可疑表现的AUROC从0.825升至0.974,敏感度从0.782升至0.935,特异度从0.759升至0.970。
- AI辅助使临床医生解读时间从274秒降至226秒,信心评分从3.90升至4.63。
- 结论认为AI辅助软件与增强识别胎儿超声中可疑先天性心脏缺陷表现相关。
核心结论
- 使用基于人工智能的软件辅助临床医生,与增强产前超声中识别可疑先天性心脏缺陷表现相关。
研究设计与样本
- 回顾性构建了来自11个中心的200例胎儿超声检查数据集,其中100例至少含一项可疑表现。
- 纳入标准为单胎妊娠、孕18-24周、患者年龄18岁及以上,且仅包含具有可解读的四腔心、左心室流出道和右心室流出道标准切面的二维灰阶动态影像。
- 7名妇产科医生和7名母胎医学专家以随机顺序分别在有无AI辅助下审阅每例检查,评估每项可疑先天性心脏缺陷表现的存在与否并给出信心评分。
主要发现
- AI辅助组识别任何可疑表现的AUROC为0.974(95%CI 0.957-0.990),未辅助组为0.825(95%CI 0.741-0.908),P=.002。
- AI辅助组敏感度为0.935(95%CI 0.892-0.978),未辅助组为0.782(95%CI 0.686-0.878)。
- AI辅助组特异度为0.970(95%CI 0.949-0.991),未辅助组为0.759(95%CI 0.630-0.887)。
- AI辅助组解读时间为226秒(95%CI 218-234),未辅助组为274秒(95%CI 265-283),P<.001。
- AI辅助组信心评分为4.63(95%CI 4.60-4.66),未辅助组为3.90(95%CI 3.85-3.95),P<.001。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: To evaluate whether artificial intelligence (AI)-based software was associated with enhanced identification of eight second-trimester fetal ultrasound findings suspicious for congenital heart defects (CHDs) among obstetrician-gynecologists (ob-gyns) and maternal-fetal medicine specialists. METHODS: A dataset of 200 fetal ultrasound examinations from 11 centers, including 100 with at least one suspicious finding, was retrospectively constituted (singleton pregnancy, 18-24 weeks of gestation, patients aged 18 years or older). Only examinations containing two-dimensional grayscale cines with interpretable four-chamber, left ventricular outflow tract, and right ventricular outflow tract standard views were included. Seven ob-gyns and seven maternal-fetal medicine specialists reviewed each examination in randomized order both with and without AI assistance and assessed the presence or absence of each finding suspicious for CHD with confidence scores. Outcomes included readers' performance in identifying the presence of any finding and each finding at the examination level, as measured by the area under the receiver operating characteristic curve (AUROC), sensitivity, and specificity. In addition, reading time and confidence were evaluated. RESULTS: The detection of any suspicious finding significantly improved for AI-aided compared with unaided readers with a significantly higher AUROC (0.974 [95% CI, 0.957-0.990] vs 0.825 [95% CI, 0.741-0.908], P =.002), sensitivity (0.935 [95% CI, 0.892-0.978] vs 0.782 [95% CI, 0.686-0.878]), and specificity (0.970 [95% CI, 0.949-0.991] vs 0.759 [95% CI, 0.630-0.887]). AI assistance also resulted in a significant decrease in clinician interpretation time and increase in clinician confidence score (226 seconds [95% CI, 218-234] vs 274 seconds [95% CI, 265-283], P <.001; 4.63 [95% CI, 4.60-4.66] vs 3.90 [95% CI, 3.85-3.95], P <.001, respectively). CONCLUSION: The use of AI-based software to assist clinicians was associated with enhanced identification of findings suspicious for CHD on prenatal ultrasonography.