生殖健康冷冻储存设施人工操作的工作量研究
该观察性时间-动作研究在4家美国IVF中心对12名操作者进行2周摄像记录,发现每次取卵需3.4次数据查询、罐平均提升1.5次、每次提升耗时11.8±9.2秒,47.25%的操作时间属疲劳性工作,各中心流程与空间容量存在差异。
要点
- 每次取卵需3.4次数据查询,罐平均提升1.5次,每次提升耗时11.8±9.2秒。
- 47.25%的冷冻储存设备操作时间属疲劳性工作。
- 各中心在液氮罐填充人员配置、频率和效率上存在差异。
核心结论
- 该时间-动作研究显示胚胎学家在人工冷冻储存流程中投入大量时间、存在任务冗余和疲劳性工作条件,不同实验室流程与空间容量存在差异,部分问题可能通过整合自动化和技术方案解决。
研究设计与样本
- 观察性时间-动作研究,由两名观察者携带数码相机在4家美国IVF中心(α、β、γ、δ)对12名执行冷冻储存任务的操作者进行为期2周的记录,并对参与者和实验室主任进行非正式访谈。
主要发现
- 平均每次从冷冻储存取卵需3.4次数据记录查询以识别一个提管,罐平均提升1.5次,每次提升平均耗时11.8±9.2秒。
- 在冷冻储存设备操作总时间中,47.25%属于疲劳性工作。
- α、β、γ中心由一人填充液氮储存罐,δ中心由两名技师协同移动和填充液氮罐,各中心填充频率、决定因素和效率不同。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
RESEARCH QUESTION: What areas of manual IVF cryostorage operations are common to the safe operation of IVF cryostorage facilities and require effort from embryologists? DESIGN: Observational time and motion data were collected by two observers equipped with the digital cameras over 2 weeks at four well-characterized US IVF centres (sites α, β, γ and δ) from 12 participants performing cryostorage tasks. To understand the work processes of the different sites and assist in the data analysis, informal interviews were conducted with the study participants and laboratory directors. Data were analysed to identify work processes that might be eliminated or diminished by automation and software improvements. RESULTS: On average, it took 3.4 data record queries per retrieval from cryostorage to identify a cane, while the canister was lifted an average of 1.5 times per retrieval, with a mean 11.8 ± 9.2 s per lift. Of the total time spent working with cryostorage equipment, 47.25% was of a fatiguing nature. Sites α, β and γ utilized one person to fill the liquid nitrogen storage Dewars, while site δ had two technicians working in tandem to move and fill the Dewars, with different frequencies and determination factors for refills and efficiencies. CONCLUSIONS: This time and motion study demonstrated significant time investment, task redundancy and fatiguing working conditions among embryologists using manual cryostorage processes. There was a disparity of processes and space capacity across different laboratories. Some of these issues may be addressed by the integration of automation and technology solutions.
引用
- Broussard A, Venier B, Rafati AS, et al. Effort in manual operation of reproductive health cryostorage facilities. Reprod Biomed Online. 2024. PMID: 39213986. doi:10.1016/j.rbmo.2024.104348
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。