聚-4-羟基丁酸酯作为耻骨后中段尿道吊带治疗压力性尿失禁的安全性与疗效报告:新型可吸收吊带的前瞻性24个月随访
前瞻性队列研究纳入17例中重度压力性尿失禁女性,行耻骨后中段尿道吊带术并植入聚-4-羟基丁酸酯(P4HB)可吸收吊带。随访24个月,报告3例严重不良事件,均可经再干预处理且无永久伤害;客观治愈率12个月为76%、24个月为47%。初步证据提示P4HB吊带可安全用于中段尿道吊带术,或可作为偏好非永久材料的患者的聚丙烯吊带替代选择。
要点
- 17例中重度压力性尿失禁女性接受P4HB可吸收吊带耻骨后中段尿道吊带术。
- 随访24个月报告3例严重不良事件,再干预后无额外或永久伤害。
- 客观治愈率12个月为76%、24个月为47%。
核心结论
- 初步证据提示P4HB吊带可安全用于中段尿道吊带术,或可作为偏好非永久材料患者的聚丙烯吊带替代选择。
研究设计与样本
- 前瞻性临床队列研究,随访24个月,在一家三级学术医院开展,纳入17例中重度压力性尿失禁女性。
主要发现
- 17例女性均顺利完成新型可吸收P4HB吊带的耻骨后中段尿道吊带术。
- 随访期间报告3例严重不良事件:术后1周无法排尿、初始吊带术后16个月因子宫内膜癌前病变行子宫切除及双侧卵巢切除、吊带网片暴露且部分装置护套残留。
- 上述参与者接受再干预,过程顺利,无额外和/或永久伤害。
- 因治愈失败而再干预包括3例再次中段尿道吊带术和1例尿道填充剂注射。
- 12个月和24个月客观治愈压力性尿失禁的患者比例分别为76%和47%。
关键数值与推荐
- 依据安全标准,至24个月终点前无需停止研究。
临床怎么用
- 原文未据此给出临床推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
STUDY OBJECTIVE: To investigate the feasibility and safety of the retropubic mid-urethral sling (MUS) procedure with a resorbable mesh implant made of poly-4-hydroxybutyrate (P4HB). DESIGN: A prospective clinical cohort study with 24 months follow-up. SETTING: A tertiary academic hospital. PATIENTS: Seventeen women with moderate to severe stress urinary incontinence (SUI). INTERVENTIONS: A retropubic MUS procedure with P4HB sling. MEASUREMENTS AND MAIN RESULTS: Seventeen women underwent an uncomplicated MUS procedure with a new sling made of resorbable P4HB. Following safety criteria, there has been no need to stop the study before 24 months' end point. Three serious adverse events were reported during follow-up: (1) inability to void urine after 1 week postsurgery, (2) a hysterectomy and bilateral oophorectomy performed due to a precancerous endometrial lesion at 16 months after initial MUS procedure and (3) exposed mesh with a part of device sheath left behind. These participants underwent an uncomplicated reintervention with no additional and/or permanent harm. Reinterventions because of failure of cure contained 3 second MUS procedures and one admission of urethral bulking agent. Seventy-six percent and 47 percent of patients were objectively cured of SUI at 12 and 24 months, respectively. CONCLUSIONS: Preliminary evidence suggests that a sling made of P4HB can be safely used for MUS procedures and could be an alternative for permanent polypropylene slings in patients that prefer the use of nonpermanent material.
引用
- van Rest KLC, Kastelein AW, Brouard KJ, et al. Safety and Efficacy Report for the Use of Poly-4-Hydroxybutyrate as a Retropubic Mid-Urethral Sling (MUS) for Stress Urinary Incontinence: A Prospective 24 Months Follow-Up of New Poly-4-Hydroxybutyrate TephaFlex SUI Bioresorbable MUS. J Minim Invasive Gynecol. 2023. PMID: 37984515. doi:10.1016/j.jmig.2023.11.013
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。