新型Fine Birth设备用于先兆早产诊断的可重复性与可用性评估

Reproducibility and usability assessment of the novel Fine Birth device for threatened preterm labor diagnosis.
PubMed 文献 · Am J Obstet Gynecol MFM · 2023-04-23

本研究在5家西班牙医院招募77例单胎孕妇,评估Fine Birth设备通过客观量化宫颈硬度诊断先兆早产的可重复性与可用性。观察者内可重复性良好(组内相关系数0.88,95%CI 0.84-0.95,Fisher检验P<.05);观察者间未达可接受值后加装侧向微型摄像头并培训,16例追加分析显示观察者间可重复性优秀(组内相关系数0.93,95%CI 0.78-0.97),干预后改善(P<.0001)。

要点

  • Fine Birth设备通过扭转波传播量化宫颈组织硬度。
  • 观察者内组内相关系数0.88,观察者间加装摄像头并培训后为0.93。
  • 设备可重复性与可用性结果良好,但临床效用仍需进一步研究。

核心结论

  • 加装侧向微型摄像头并培训后,Fine Birth设备获得稳健的可重复性与可用性结果,有望客观量化宫颈硬度、诊断先兆早产并预测自发性早产风险。
  • 尚需进一步研究以证明该设备的临床效用。

研究设计与样本

  • 在5家西班牙医院的妇产科随访期间招募77例单胎孕妇。
  • 纳入标准为年龄≥18岁、胎儿正常且妊娠无并发症、无胎膜膨出、子宫畸形、既往宫颈手术或乳胶过敏,并签署书面知情同意。
  • 使用Fine Birth设备评估宫颈组织硬度,每例由2名不同操作者获取2次有效测量。
  • 采用组内相关系数(95%CI)和Fisher检验P值评估观察者内与观察者间可重复性,并根据临床医生与参与者反馈评估可用性。

主要发现

  • 观察者内可重复性良好(组内相关系数0.88,95%CI 0.84-0.95,Fisher检验P<.05)。
  • 观察者间可重复性未达可接受值(组内相关系数<0.75),随后在阴道探头加装侧向微型摄像头并对操作者培训。
  • 16例追加受试者分析显示观察者间可重复性优秀(组内相关系数0.93,95%CI 0.78-0.97),干预后改善(P<.0001)。

局限性

  • 作者指出尚需进一步研究以证明该设备的临床效用。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Preterm delivery is considered the leading cause of mortality worldwide in children under 5 years old. Approximately 45 million pregnant women are hospitalized yearly for threatened preterm labor. However, only 50% of pregnancies complicated by threatened preterm labor end in delivery before the estimated date, classifying the rest as false threatened preterm labor. The ability of current diagnostic methods to predict threatened preterm labor is low (low positive predictive value), ranging between 8% and 30%. This highlights the need for a solution that accurately detects and differentiates between false and real threatened preterm labors in women who attend obstetrical clinics and hospital emergency departments with delivery symptoms. OBJECTIVE: Primarily, this aimed to assess the reproducibility and usability of a novel medical device, the Fine Birth, aimed at accurately diagnosing threatened preterm labor through the objective quantification of pregnant women's cervical consistency. Secondarily, this study aimed to evaluate the effect of training and the incorporation of a lateral microcamera on the device's reliability and usability outcomes. STUDY DESIGN: A total of 77 singleton pregnant women were recruited during their follow-up visits to the obstetrical and gynecologic departments at 5 Spanish hospitals. The eligibility criteria included pregnant women aged ≥18 years; women with a normal fetus and uncomplicated pregnancy; women without prolapse of membranes, uterine anomalies, previous cervical surgery, or latex allergy; and women signing the informed written consent. Cervical tissue stiffness was assessed using the Fine Birth device, whose technology is based on the propagation of torsional waves through the studied tissue. Cervical consistency measurements were taken for each woman until obtaining 2 valid measurements by 2 different operators. The intraobserver and interobserver reproducibilities of the Fine Birth measurements were assessed using the intraclass correlation coefficients with a 95% confidence interval and the Fisher test P value. The usability was evaluated on the basis of the clinicians' and participants' feedback. RESULTS: There was good intraobserver reproducibility (intraclass correlation coefficient, 0.88; 95% confidence interval, 0.84-0.95; Fisher test P value<.05). As the results obtained for the interobserver reproducibility did not reach the desired acceptable values (intraclass correlation coefficient of <0.75), a lateral microcamera was added to the Fine Birth intravaginal probe, and the operators involved in the clinical investigation received the corresponding training with the modified device. The analysis of 16 additional subjects demonstrated excellent interobserver reproducibility (intraclass correlation coefficient, 0.93; 95% confidence interval, 0.78-0.97) and an improvement after the intervention (P<.0001). CONCLUSION: The robust reproducibility and usability results obtained after the insertion of a lateral microcamera and the corresponding training make the Fine Birth a promising novel device to objectively quantify the patient's cervical consistency, diagnose threatened preterm labor, and, thus, predict the risk of spontaneous preterm birth. Further research is needed to demonstrate the clinical utility of the device.