运动校正胎儿身体磁共振成像在正常与异常胎儿中提供可靠的3D肺体积
单中心胎儿医学转诊中心的初步研究,纳入16例胎龄22–32周的胎儿MRI数据集(先天性膈疝6例、先天性肺气道畸形2例、健康对照8例),采用形变切片到容积配准(DSVR)重建胎儿身体3D各向同性(0.85 mm)容积并行半自动肺分割,将3D总肺体积(TLV)与传统的2D体积测量进行比较,异常病例参照由100例正常胎儿生成的常模。
要点
- DSVR导出的TLV与2D测量高度相关,但体积持续偏低,偏倚-1.44 cm3(95%限:-2.6至-0.3)。
- DSVR分辨率提高,即使在运动伪影或肺体积很低时也能排除肺门结构。
- DSVR有助于分析运动伪影扫描,且不存在2D分割固有的插值误差,可能改善咨询与手术规划。
核心结论
- DSVR用于胎儿肺MRI有助于分析运动伪影扫描,且不存在2D分割固有的插值误差;它增加了所采集数据在3D空间显示器官和体积定量方面的信息含量,可能改善咨询与手术规划。
研究设计与样本
- 单中心胎儿医学转诊中心的初步研究,纳入16例胎龄22–32周的胎儿MRI数据集,诊断为先天性膈疝(n=6)、先天性肺气道畸形(n=2)和健康对照(n=8)。
- 采用形变切片到容积配准(DSVR)重建胎儿身体3D各向同性(0.85 mm)容积,随后行半自动肺分割。
- 将3D TLV与传统的2D体积测量进行比较;异常病例参照由100例正常胎儿生成的常模,该常模的TLV仅由DSVR计算。
主要发现
- DSVR导出的TLV值与2D测量高度相关,但体积持续偏低,偏倚-1.44 cm3(95%限:-2.6至-0.3)。
- DSVR分辨率提高,即使在运动伪影或肺体积很低时也能排除肺门结构。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: To calculate 3D-segmented total lung volume (TLV) in fetuses with thoracic anomalies using deformable slice-to-volume registration (DSVR) with comparison to 2D-manual segmentation. To establish a normogram of TLV calculated by DSVR in healthy control fetuses. METHODS: A pilot study at a single regional fetal medicine referral centre included 16 magnetic resonance imaging (MRI) datasets of fetuses (22-32 weeks gestational age). Diagnosis was CDH (n = 6), CPAM (n = 2), and healthy controls (n = 8). Deformable slice-to-volume registration was used for reconstruction of 3D isotropic (0.85 mm) volumes of the fetal body followed by semi-automated lung segmentation. 3D TLV were compared to traditional 2D-based volumetry. Abnormal cases referenced to a normogram produced from 100 normal fetuses whose TLV was calculated by DSVR only. RESULTS: Deformable slice-to-volume registration-derived TLV values have high correlation with the 2D-based measurements but with a consistently lower volume; bias -1.44 cm3 [95% limits: -2.6 to -0.3] with improved resolution to exclude hilar structures even in cases of motion corruption or very low lung volumes. CONCLUSIONS: Deformable slice-to-volume registration for fetal lung MRI aids analysis of motion corrupted scans and does not suffer from the interpolation error inherent to 2D-segmentation. It increases information content of acquired data in terms of visualising organs in 3D space and quantification of volumes, which may improve counselling and surgical planning.
引用
- Davidson J, Uus A, Egloff A, et al. Motion corrected fetal body magnetic resonance imaging provides reliable 3D lung volumes in normal and abnormal fetuses. Prenat Diagn. 2022. PMID: 35262959. doi:10.1002/pd.6129
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。