本研究开发并验证了产前超声人工智能诊断系统PAICS,用于在标准超声参考切面中检测胎儿颅内异常模式以筛查先天性中枢神经系统畸形。系统基于YOLOv3实时卷积神经网络算法,使用来自中国两家三级医院的图像进行训练、微调和内部验证,并用第三家三级医院的图像进行外部验证,同时与不同经验水平的超声医师比较。PAICS在识别10种颅内图像模式方面表现优异,诊断性能与专家超声医师相当,但所需时间显著更少。

要点

  • PAICS在内部验证图像数据集中的宏平均AUC为0.933,微平均AUC为0.977。
  • PAICS在实时扫查场景中的宏平均AUC为0.969,微平均AUC为0.981。
  • PAICS诊断性能与专家超声医师相当,但每张图像所需时间显著更少(0.025秒 vs 4.4秒,P < 0.001)。

核心结论

  • 在图像数据集和实时扫查场景中,PAICS对多种胎儿中枢神经系统异常均达到优异的诊断性能。
  • PAICS的性能与专家超声医师相当,但所需时间更少。
  • 卷积神经网络算法可经训练用于检测胎儿中枢神经系统异常。
  • PAICS有潜力成为临床实践中筛查胎儿中枢神经系统畸形的有效且高效的工具。

研究设计与样本

  • 研究从中国两家三级医院的数据库中获取18-40孕周正常胎儿及中枢神经系统畸形胎儿的神经超声图像。
  • 图像按8:1:1比例随机分配至训练集、微调集和内部验证集,用于开发和评估PAICS。
  • 系统基于实时卷积神经网络算法YOLOv3构建。
  • 使用第三家三级医院的图像数据集对PAICS进行外部验证,并与不同经验水平的超声医师比较性能。
  • 采用前瞻性视频数据集评估PAICS在实时扫查场景中的性能。

主要发现

  • 共使用来自16 297例妊娠的43 890张图像和来自166例妊娠的169段视频开发和验证PAICS。
  • 内部验证图像数据集中,PAICS识别10种颅内图像模式的宏平均AUC为0.933(95% CI,0.798-1.000),微平均AUC为0.977(95% CI,0.970-0.985)。
  • 外部验证图像数据集中,宏平均AUC为0.902(95% CI,0.816-0.989),微平均AUC为0.898(95% CI,0.885-0.911)。
  • 实时扫查场景中,宏平均AUC为0.969(95% CI,0.886-1.000),微平均AUC为0.981(95% CI,0.974-0.988)。
  • PAICS与专家超声医师在宏平均和微平均准确率(P = 0.863和P = 0.775)、敏感度(P = 0.883,P = 0.846)和AUC(P = 0.891,P = 0.788)方面表现相当,但每张图像所需时间显著更少(PAICS 0.025秒 vs 专家4.4秒,P < 0.001)。

临床怎么用

  • PAICS有潜力成为临床实践中筛查胎儿中枢神经系统畸形的有效且高效的工具。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: To develop and validate an artificial intelligence system, the Prenatal ultrasound diagnosis Artificial Intelligence Conduct System (PAICS), to detect different patterns of fetal intracranial abnormality in standard sonographic reference planes for screening for congenital central nervous system (CNS) malformations. METHODS: Neurosonographic images from normal fetuses and fetuses with CNS malformations at 18-40 gestational weeks were retrieved from the databases of two tertiary hospitals in China and assigned randomly (ratio, 8:1:1) to training, fine-tuning and internal validation datasets to develop and evaluate the PAICS. The system was built based on a real-time convolutional neural network (CNN) algorithm, You Only Look Once, version 3 (YOLOv3). An image dataset from a third tertiary hospital was used to further validate, externally, the performance of the PAICS and to compare its performance with that of sonologists with different levels of expertise. Furthermore, a prospective video dataset was employed to evaluate the performance of the PAICS in a real-time scan scenario. The diagnostic accuracy, sensitivity, specificity and area under the receiver-operating-characteristics curve (AUC) were calculated to assess the performance of the PAICS and to compare this with the performance of sonologists with different levels of experience. RESULTS: In total, 43 890 images from 16 297 pregnancies and 169 videos from 166 pregnancies were used to develop and validate the PAICS. The system achieved excellent performance in identifying 10 types of intracranial image pattern, with macro- and microaverage AUCs, respectively, of 0.933 (95% CI, 0.798-1.000) and 0.977 (95% CI, 0.970-0.985) for the internal validation image dataset, 0.902 (95% CI, 0.816-0.989) and 0.898 (95% CI, 0.885-0.911) for the external validation image dataset and 0.969 (95% CI, 0.886-1.000) and 0.981 (95% CI, 0.974-0.988) in the real-time scan setting. The performance of the PAICS was comparable to that of expert sonologists in terms of macro- and microaverage accuracy (P = 0.863 and P = 0.775, respectively), sensitivity (P = 0.883, P = 0.846) and AUC (P = 0.891, P = 0.788), but required significantly less time (0.025 s per image for PAICS vs 4.4 s for experts, P < 0.001). CONCLUSIONS: Both in the image dataset and in the real-time scan setting, the PAICS achieved excellent diagnostic performance for various fetal CNS abnormalities. Its performance was comparable to that of experts, but it required less time. A CNN algorithm can be trained to detect fetal CNS abnormalities. The PAICS has the potential to be an effective and efficient tool in screening for fetal CNS malformations in clinical practice. © 2021 International Society of Ultrasound in Obstetrics and Gynecology.