将个人设备床旁超声纳入产科保健:一项验证研究
本研究在城市产前超声诊断中心前瞻性观察19-39周孕妇,由专家超声医师分别用标准超声机和个人设备床旁超声(P-POCUS)评估估计胎儿体重。共完成100对扫描,组内相关系数0.99(P<.0001),Bland-Altman平均差异53克,95%一致性界限-178至283克,Pearson相关r=0.99(P<.0001)。结论认为P-POCUS是可靠的基本产科超声工具,可能增加全球产科超声可及性。
要点
- 个人设备床旁超声与标准超声机测估计胎儿体重的组内相关系数为0.99(P<.0001)。
- Bland-Altman分析平均差异53克,95%一致性界限-178至283克。
- P-POCUS是可靠的基本产科超声工具,可能增加全球产科超声可及性。
核心结论
- 个人设备床旁超声设备是执行基本产科超声的可靠工具,并可能增加全球产科超声的可及性。
研究设计与样本
- 这是一项前瞻性观察性研究,纳入城市产前超声诊断中心19-39周妊娠患者,由专家超声医师分别使用标准超声机和个人设备床旁超声评估估计胎儿体重。
- 2020年10月至2020年12月共完成100对扫描。
- 通过组内相关系数、Bland-Altman图和Pearson相关系数评估估计胎儿体重的统计信度和一致性。
主要发现
- 估计胎儿体重的组内相关系数为0.99(P<.0001),一致性接近完美。
- Bland-Altman分析显示平均差异为53克,95%一致性界限为-178克至283克。
- Pearson相关显示测量之间接近完美相关(r=0.99,P<.0001)。
临床怎么用
- 原文结论认为个人设备床旁超声设备是执行基本产科超声的可靠工具,并可能增加全球产科超声的可及性。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Personal-device-based point-of-care-ultrasound (P-POCUS) probes plug directly into a cell phone or tablet to function as its display, creating the potential to increase access to obstetric ultrasonography in complex healthcare settings (COVID units, low resource settings); however, new technology must be proven to be reliable in the obstetric setting before integrating into practice. OBJECTIVE: To evaluate the intraclass correlation (reliability) of personal-device-based-point-of-care-ultrasound devices as compared with standard ultrasound machines in obstetrics. STUDY DESIGN: This was a prospective, observational study of patients between 19-39 weeks gestation in an urban, prenatal ultrasound diagnosis center. Each patient underwent assessment by an expert sonographer using standard ultrasound machines and personal-device-based-point-of-care-ultrasound devices to determine estimated fetal weight. The statistical reliability and agreement between the estimated fetal weights was assessed through intraclass correlation coefficients, Bland-Altman plots, and Pearson correlation coefficients. RESULTS: 100 paired sets of scans were performed from October 2020 to December 2020. For the estimated fetal weights, there was near-perfect agreement, with an intraclass correlation coefficient of 0.99 (P<.0001). Bland-Altman analysis showed an average difference of 53 grams, with 95% limit of agreement between -178 grams and 283 grams. Pearson correlation showed near-perfect correlation between the measurements (r=0.99, P<.0001). CONCLUSION: personal-device-based point-of-care-ultrasound devices are reliable tools for performing basic obstetrical ultrasound and have the potential to increase access to obstetrical ultrasound worldwide.
引用
- Leggett CB, Naqvi M, Esakoff TF, et al. Incorporating personal-device-based point-of-care ultrasound into obstetric care: a validation study. Am J Obstet Gynecol. 2021. PMID: 34774825. doi:10.1016/j.ajog.2021.11.031
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。